[螺旋断层放射治疗中兆伏级计算机断层扫描成像剂量的测量与分析]

[Measurement and analysis of the imaging dose with megavoltage computed tomography for helical tomotherapy].

作者信息

Xu Shou-Ping, Xie Chuan-Bin, Ju Zhong-Jian, Dai Xiang-Kun, Gong Han-Shun, Guo Yan-Yan, Wang Lian-Yuan

机构信息

Department of Radiation Oncology, PLA General Hospital, Beijing, 100853, P.R. China.

出版信息

Ai Zheng. 2009 Aug;28(8):886-9. doi: 10.5732/cjc.008.10632.

Abstract

BACKGROUND & OBJECTIVE: During the helical tomotherapy process, megavoltage computed tomography (MVCT) images are usually used for guiding the precise setup of patients before/after treatment delivery, which would certainly increase the total dose for patients. This study was to investigate the imaging dose of MVCT using the body and head phantom on a tomotherapy machine.

METHODS

A set of cylindrical body and head phantoms was adopted for scanning with different pitch values (1.0/2.0/3.0), lengths (4.8/7.2/9.6/12/14.4 cm) and patient setups on the couch of tomotherapy system. The average MVCT imaging doses were measured using A1SL chambers inserted in the phantoms with preset geometry. The dose uniformity was qualitatively analyzed.

RESULTS

The MVCT scanning dose for the body phantom was between 0.599 and 2.876 cGy during each treatment delivery, while the dose for the head phantom was between 0.913 and 3.231 cGy. Two major parameters, the assigned pitch numbers and scanning lengths, were the most important impacts to the dose variation. The MVCT dose was inversely proportional to the CT pitch value. With respect to the scanning length, the doses responded differently along the radial direction of the phantoms with different setup criteria.

CONCLUSION

The results may provide a reliable guidance for proper planning design of the scanning region, which is valuable to help minimize the extra doses to patient.

摘要

背景与目的

在螺旋断层放射治疗过程中,兆伏级计算机断层扫描(MVCT)图像通常用于在治疗前/后引导患者的精确摆位,这无疑会增加患者的总剂量。本研究旨在使用断层放射治疗机上的体模和头模研究MVCT的成像剂量。

方法

采用一组圆柱形体模和头模,在断层放射治疗系统的治疗床上以不同的螺距值(1.0/2.0/3.0)、长度(4.8/7.2/9.6/12/14.4 cm)和患者摆位进行扫描。使用插入具有预设几何形状体模中的A1SL电离室测量平均MVCT成像剂量。对剂量均匀性进行定性分析。

结果

每次治疗时,体模的MVCT扫描剂量在0.599至2.876 cGy之间,而头模的剂量在0.913至3.231 cGy之间。两个主要参数,即指定的螺距数和扫描长度,对剂量变化影响最大。MVCT剂量与CT螺距值成反比。关于扫描长度,在不同摆位标准下,沿体模径向的剂量反应不同。

结论

研究结果可为扫描区域的合理规划设计提供可靠指导,这对于帮助尽量减少患者的额外剂量具有重要价值。