使用Presage/光学CT对放射物理中心头颈IMRT体模中相对三维剂量学的可行性进行研究。

Investigation of the feasibility of relative 3D dosimetry in the Radiologic Physics Center Head and Neck IMRT phantom using presage/optical-CT.

作者信息

Sakhalkar Harshad, Sterling David, Adamovics John, Ibbott Geoffrey, Oldham Mark

机构信息

Department of Radiation Oncology Physics, Duke University Medical Center, Durham, North Carolina 27710, USA.

出版信息

Med Phys. 2009 Jul;36(7):3371-7. doi: 10.1118/1.3148534.

Abstract

This study presents the application of the Presage/optical-CT 3D dosimetry system for relative dosimetry in the Radiologic Physics Center (RPC) Head and Neck (H&N) IMRT phantom. Performance of the system was evaluated by comparison with the "gold-standard" RPC credentialing test. A modified Presage cylindrical insert was created that extended the capability of the RPC H&N phantom to 3D dosimetry. The RPC phantom was taken through the entire treatment planning procedure with both the standard RPC insert and the modified Presage insert. An IMRT plan was created to match the desired dose constraints of the credentialing test. This plan was delivered twice to the RPC phantom: first containing the standard insert, and then again containing the Presage insert. After irradiation, the standard insert was sent for routine credentialing analysis; including point dose measurements (TLD) and planar Gafchromic EBT film measurement. The 3D dose distribution from Presage was read out at Duke using the OCTOPUS 5X optical-CT scanner. The Presage distribution was compared with gold-standard EBT measurement (determined by the RPC) and the calculated Eclipse distribution. The agreement between the normalized EBT, Presage, and Eclipse distributions, in the central axial plane was evaluated using profiles and gamma-map comparisons (4% dose difference and 3 mm distance to agreement). Profiles showed good agreement between EBT, Presage, and Eclipse distributions. 2D gamma-map comparisons between all three modalities showed at least 98% pass rate. The excellent agreement between Presage and EBT in the central plane established Presage as a standard against which to evaluate the accuracy of the 3D calculated Eclipse distribution. A gamma comparison between normalized Presage and Eclipse 3D distributions gave an overall pass rate of approximately 94%. In conclusion, the Presage/optical-CT system was found to be feasible for relative 3D dosimetry in the RPC IMRT H&N phantom. The potential to extend the RPC IMRT credentialing procedure to 3D may be feasible provided accurate calibration to dose (Gy) and robustness to shipping stress are demonstrated.

摘要

本研究介绍了Presage/光学CT 3D剂量测定系统在放射物理中心(RPC)头颈部(H&N)调强放疗(IMRT)体模中进行相对剂量测定的应用。通过与“金标准”RPC认证测试进行比较,评估了该系统的性能。制作了一种改良的Presage圆柱形插入物,将RPC H&N体模的功能扩展到3D剂量测定。使用标准RPC插入物和改良的Presage插入物,使RPC体模经历整个治疗计划过程。创建了一个IMRT计划,以符合认证测试的预期剂量限制。该计划分两次照射RPC体模:第一次包含标准插入物,第二次包含Presage插入物。照射后,将标准插入物送去进行常规认证分析;包括点剂量测量(热释光剂量计)和平面Gafchromic EBT胶片测量。在杜克大学使用OCTOPUS 5X光学CT扫描仪读取Presage的3D剂量分布。将Presage分布与金标准EBT测量结果(由RPC确定)以及计算得到的Eclipse分布进行比较。使用剖面和伽马图比较(4%剂量差异和3毫米一致距离)评估归一化EBT、Presage和Eclipse分布在中心轴平面上的一致性。剖面显示EBT、Presage和Eclipse分布之间具有良好的一致性。所有三种模态之间的二维伽马图比较显示通过率至少为98%。Presage和EBT在中心平面上的出色一致性确立了Presage作为评估3D计算Eclipse分布准确性的标准。归一化Presage和Eclipse 3D分布之间的伽马比较给出的总体通过率约为94%。总之,发现Presage/光学CT系统在RPC IMRT H&N体模中进行相对3D剂量测定是可行的。如果能证明对剂量(Gy)的准确校准以及对运输应力的稳健性,将RPC IMRT认证程序扩展到3D的可能性可能是可行的。

相似文献

2
Investigating the Feasibility of 3D Dosimetry in the RPC IMRT H&N Phantom.
J Phys Conf Ser. 2009;164(2009):12058. doi: 10.1088/1742-6596/164/1/012058.
4
A comprehensive evaluation of the PRESAGE/optical-CT 3D dosimetry system.
Med Phys. 2009 Jan;36(1):71-82. doi: 10.1118/1.3005609.
5
A practical three-dimensional dosimetry system for radiation therapy.
Med Phys. 2006 Oct;33(10):3962-72. doi: 10.1118/1.2349686.
7
Investigation into the feasibility of using PRESAGE/optical-CT dosimetry for the verification of gating treatments.
Phys Med Biol. 2010 Apr 21;55(8):2187-201. doi: 10.1088/0031-9155/55/8/005. Epub 2010 Mar 26.
8
Study of PRESAGE® dosimeter for end-to-end 3D radiotherapy verification using an anthropomorphic phantom with bespoke dosimeter insert.
Phys Eng Sci Med. 2024 Sep;47(3):955-966. doi: 10.1007/s13246-024-01418-9. Epub 2024 Apr 18.
10
SU-E-T-95: Investigation of 3D Dosimetry for an Anthropomorphic Spine Phantom.
Med Phys. 2012 Jun;39(6Part11):3724. doi: 10.1118/1.4735152.

引用本文的文献

1
A Review of PRESAGE Radiochromic Polymer and the Compositions for Application in Radiotherapy Dosimetry.
Polymers (Basel). 2022 Jul 16;14(14):2887. doi: 10.3390/polym14142887.
2
Novel methodologies for dosimetry audits: Adapting to advanced radiotherapy techniques.
Phys Imaging Radiat Oncol. 2018 Mar 19;5:76-84. doi: 10.1016/j.phro.2018.03.002. eCollection 2018 Jan.
3
Dosimetric characterization of a body-conforming radiochromic sheet.
J Appl Clin Med Phys. 2020 Mar;21(3):167-177. doi: 10.1002/acm2.12838. Epub 2020 Feb 26.
5
The effect of motion on IMRT - looking at interplay with 3D measurements.
J Phys Conf Ser. 2013;444:012049. doi: 10.1088/1742-6596/444/1/012049.
6
An investigation of PRESAGE® 3D dosimetry for IMRT and VMAT radiation therapy treatment verification.
Phys Med Biol. 2015 Mar 21;60(6):2217-30. doi: 10.1088/0031-9155/60/6/2217. Epub 2015 Feb 16.
7
On the feasibility of comprehensive high-resolution 3D remote dosimetry.
Med Phys. 2014 Jul;41(7):071706. doi: 10.1118/1.4884018.
9
Comprehensive quality assurance for base of skull IMRT.
J Phys Conf Ser. 2013;4444(1):012050. doi: 10.1088/1742-6596/444/1/012050.

本文引用的文献

1
A comprehensive evaluation of the PRESAGE/optical-CT 3D dosimetry system.
Med Phys. 2009 Jan;36(1):71-82. doi: 10.1118/1.3005609.
2
Radiological properties of the PRESAGE and PAGAT polymer dosimeters.
Appl Radiat Isot. 2008 Dec;66(12):1970-4. doi: 10.1016/j.apradiso.2008.06.005. Epub 2008 Jun 11.
4
Challenges in credentialing institutions and participants in advanced technology multi-institutional clinical trials.
Int J Radiat Oncol Biol Phys. 2008;71(1 Suppl):S71-5. doi: 10.1016/j.ijrobp.2007.08.083.
7
A practical three-dimensional dosimetry system for radiation therapy.
Med Phys. 2006 Oct;33(10):3962-72. doi: 10.1118/1.2349686.
8
Characterisation of PRESAGE: A new 3-D radiochromic solid polymer dosemeter for ionising radiation.
Radiat Prot Dosimetry. 2006;120(1-4):107-12. doi: 10.1093/rpd/nci555. Epub 2006 Jun 17.
10
Characterization of a new radiochromic three-dimensional dosimeter.
Med Phys. 2006 May;33(5):1338-45. doi: 10.1118/1.2192888.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验