Kitthaweesin Kitthisak, Mungsing Warataporn
Department of Ophthalmology, Faculty of Medicine, Khon Kaen University, Khon Kaen, Thailand.
J Med Assoc Thai. 2009 Aug;92(8):1046-9.
To investigate the agreement and reproducibility of contact and immersion ultrasound biometry in an axial length measurement and intraocular lens power calculation.
Prospective cross-sectional study.
Axial length was measured in 198 cataract patients using both contact and immersion techniques, randomly assigned to two measurers, one with more experience than the other Both techniques were repeated by each measurer. The mean difference of both techniques was calculated for the difference of the intraocular lens power.
The respective mean differences of both techniques for the first and the second measurers were 0.03 and 0.07 mm. The repeatability coefficient was two-fold for the contact but was similar for the immersion technique. The respective difference of the intraocular lens power of both measurers was 1.15 and 1.65 diopters (D). A statistically significant agreement was noted between both techniques. The reproducibility of both techniques was similar when performed by an experienced measurer however, the less experienced measurer had greater reproducibility with the immersion technique.
The difference in the intraocular lens power from both techniques may have clinical significance; hence, the immersion technique should be considered, particularly for less-experienced measurer.
研究接触式和浸没式超声生物测量法在眼轴长度测量及人工晶状体屈光力计算中的一致性和可重复性。
前瞻性横断面研究。
对198例白内障患者使用接触式和浸没式技术测量眼轴长度,患者随机分配给两名测量者,其中一名经验比另一名丰富。每位测量者对两种技术均重复测量。计算两种技术在人工晶状体屈光力差异方面的平均差值。
第一位测量者两种技术各自的平均差值为0.03mm,第二位测量者为0.07mm。接触式技术的重复性系数为两倍,但浸没式技术的重复性系数相似。两名测量者人工晶状体屈光力各自的差值分别为1.15和1.65屈光度(D)。两种技术之间存在统计学上的显著一致性。由经验丰富的测量者操作时,两种技术的可重复性相似;然而,经验较少的测量者使用浸没式技术时可重复性更高。
两种技术在人工晶状体屈光力方面的差异可能具有临床意义;因此,应考虑采用浸没式技术,尤其是对于经验较少的测量者。