Zhang Liangyi, Reilly James P
Department of Chemistry, Indiana University, 800 East Kirkwood Avenue, Bloomington, Indiana 47405, USA.
Anal Chem. 2009 Sep 15;81(18):7829-38. doi: 10.1021/ac9012557.
Photodissociation with 157 nm light was implemented in an ABI model 4700 matrix-assisted laser desorption ionization (MALDI) tandem time-of-flight (TOF) mass spectrometer for peptide analysis. With a homemade computer program to control the light timing based on the m/z of each precursor ion, the photodissociation setup was seamlessly automated with the mass spectrometer. Peptide photodissociation in this apparatus yielded fragments similar to those observed in previous experiments with a home-built tandem-TOF mass spectrometer. Peptides having arginine at their C-termini yielded high-energy x-, v-, and w- type fragments, while peptides with N-terminal arginine produced many a- and d- type ions. Abundant immonium ions were also generated. High-quality photodissociation spectra were obtained with as little as 5 fmol of peptides. In the analysis of various tryptic peptides, photodissociation provided much more sequence information than the conventional TOF-TOF collision induced dissociation (CID). Because of the high fragmentation efficiency, sensitivity was not sacrificed to achieve this.
在一台ABI 4700型基质辅助激光解吸电离(MALDI)串联飞行时间(TOF)质谱仪中采用157 nm光进行光解离,用于肽分析。通过一个自制的计算机程序,根据每个前体离子的m/z来控制光的时间,光解离装置与质谱仪实现了无缝自动化。该装置中的肽光解离产生的碎片与之前使用自制串联TOF质谱仪的实验中观察到的碎片相似。C端含有精氨酸的肽产生高能x、v和w型碎片,而N端含有精氨酸的肽产生许多a和d型离子。还产生了大量的亚铵离子。使用低至5 fmol的肽即可获得高质量的光解离光谱。在对各种胰蛋白酶肽的分析中,与传统的TOF-TOF碰撞诱导解离(CID)相比,光解离提供了更多的序列信息。由于具有高碎片化效率,实现这一点并未牺牲灵敏度。