US Food and Drug Administration, Center for Devices and Radiological Health, OSEL/DB, Silver Spring, MD 20993, USA.
Talanta. 2009 Dec 15;80(2):1016-9. doi: 10.1016/j.talanta.2009.07.031. Epub 2009 Jul 23.
The objective of this study was to establish a reasonably simple and reliable method to measure very low concentrations of polyhexamethylene biguanide (PHMB) in multipurpose contact lens solutions (MPSs). By using a weak cation exchange solid phase extraction cartridge to extract the PHMB from MPS, followed by HPLC analysis using an evaporative light scattering detector, low levels (0.1 ppm) of PHMB were detected. Application of this method to a series of off-the-shelf MPS with PHMB as the active ingredient demonstrated these solutions contain 1 ppm. The contact lens solution with hydrogen peroxide as the active ingredient gave no peak where the PHMB peak eluted. The Polyquad contact lens solution generated a peak close to the retention time of PHMB. Recovery of PHMB from fortified hydrogen peroxide contact lens solution was good at 0.25 ppm and above; 105% with a RSD of 17% or less. The repeatability of the HPLC system ranged from 4 to 11% RSD; the reproducibility of the entire method was less than 17.5% RSD. Storage and stability studies indicated that storage of MPS with PHMB for chemical analysis are not temperature dependent, but are affected by the composition of the container in which the contact lens solution is stored.
本研究旨在建立一种合理简单且可靠的方法来测定多功能隐形眼镜护理液(MPS)中聚六亚甲基双胍(PHMB)的极低浓度。通过使用弱阳离子交换固相萃取小柱从 MPS 中提取 PHMB,然后使用蒸发光散射检测器进行 HPLC 分析,检测到低浓度(0.1ppm)的 PHMB。将该方法应用于一系列含有 PHMB 作为活性成分的市售 MPS 中,结果表明这些溶液中含有 1ppm 的 PHMB。含有过氧化氢作为活性成分的隐形眼镜溶液在 PHMB 洗脱的位置没有峰出现。Polyquad 隐形眼镜溶液在接近 PHMB 保留时间的位置产生一个峰。从加标过的过氧化氢隐形眼镜溶液中回收 PHMB 的效果良好,在 0.25ppm 及以上时回收率为 105%,相对标准偏差(RSD)小于 17%。HPLC 系统的重复性范围为 4%至 11%RSD;整个方法的重现性小于 17.5%RSD。储存和稳定性研究表明,含有 PHMB 的 MPS 进行化学分析时不受温度影响,但受储存隐形眼镜溶液的容器成分影响。