常规和超级光滑弹性结扎线在口腔内使用后的物理性能。

Physical properties of conventional and Super Slick elastomeric ligatures after intraoral use.

机构信息

Department of Orthodontics, Charles Clifford Dental Hospital, South Yorkshire, Sheffield, UK.

出版信息

Angle Orthod. 2010 Jan;80(1):175-81. doi: 10.2319/013009-63.1.

Abstract

OBJECTIVE

To investigate the change in the physical properties of conventional and Super Slick elastomeric ligatures after they have been in the mouth.

MATERIALS AND METHODS

Nine healthy volunteers took part. One orthodontic bracket was bonded to a premolar tooth in each of the four quadrants of the mouth. Two conventional and two Super Slick elastomeric ligatures were placed at random locations on either side of the mouth. The ligatures were collected after various time intervals and tested using an Instron Universal testing machine. The two outcome measures were failure load and the static frictional resistance.

RESULTS

The failure load for conventional ligatures was reduced to 67% of the original value after 6 weeks in situ. Super Slick elastomeric ligatures showed a comparable reduction after 6 weeks in situ (63% of original value). There were no statistical differences in the static friction between conventional and Super Slick elastomerics that had been in situ for either 24 hours (P = .686) or 6 weeks (P = .416). There was a good correlation between failure load and static friction (r = .49).

CONCLUSIONS

There were statistically significant differences in the failure loads of elastomerics that had not be placed in the mouth and those that had been in the mouth for 6 weeks. There were no differences in the static frictional forces produced by conventional and Super Slick ligatures either before or after they had been placed in the mouth. There appears to be a direct proportional relationship between failure load and static friction of elastomeric ligatures.

摘要

目的

研究传统和超级光滑弹性结扎线在口腔内使用后的物理性能变化。

材料与方法

共有 9 名健康志愿者参与研究。在每个志愿者口腔的四个象限中的每一个磨牙上粘结一个正畸托槽。将两条传统弹性结扎线和两条超级光滑弹性结扎线随机放置在口腔两侧的任意位置。结扎线在不同的时间间隔后收集并使用 In stron 万能试验机进行测试。两个结果测量指标是失效负载和静态摩擦阻力。

结果

在体内放置 6 周后,传统结扎线的失效负载降低至原始值的 67%。超级光滑弹性结扎线在体内放置 6 周后也表现出类似的降低(原始值的 63%)。在体内放置 24 小时(P =.686)或 6 周(P =.416)时,传统和超级光滑弹性体之间的静态摩擦力无统计学差异。失效负载与静态摩擦力之间存在良好的相关性(r =.49)。

结论

未放入口腔的弹性体的失效负载与放入口腔 6 周后的弹性体的失效负载有统计学显著差异。在放入口腔前后,传统和超级光滑结扎线产生的静态摩擦力没有差异。弹性结扎线的失效负载和静态摩擦力之间似乎存在直接的比例关系。

相似文献

3
Friction of orthodontic elastomeric ligatures with different dimensions.不同尺寸正畸弹性结扎丝的摩擦力
Angle Orthod. 2005 May;75(3):421-5. doi: 10.1043/0003-3219(2005)75[421:FOOELW]2.0.CO;2.
9
The effect of ligation method on friction in sliding mechanics.结扎方法对滑动力学中摩擦力的影响。
Am J Orthod Dentofacial Orthop. 2003 Apr;123(4):416-22. doi: 10.1067/mod.2003.14.

引用本文的文献

8
Tensile properties of orthodontic elastomeric chains.正畸弹性链的拉伸性能。
J Orofac Orthop. 2010 Sep;71(5):330-8. doi: 10.1007/s00056-010-1020-x. Epub 2010 Oct 21.

本文引用的文献

1
Sliding resistance with esthetic ligatures: an in-vitro study.美学结扎丝的滑动阻力:一项体外研究。
Am J Orthod Dentofacial Orthop. 2008 Mar;133(3):340.e1-7. doi: 10.1016/j.ajodo.2007.08.015.
3
A comparison of different ligation methods on friction.不同结扎方法对摩擦力影响的比较。
Am J Orthod Dentofacial Orthop. 2006 Nov;130(5):666-70. doi: 10.1016/j.ajodo.2006.04.021.
5
Resistance to sliding with 3 types of elastomeric modules.使用3种弹性体组件时的抗滑动性能
Am J Orthod Dentofacial Orthop. 2005 Jun;127(6):670-5; quiz 754. doi: 10.1016/j.ajodo.2004.01.025.
8
Tensile properties of orthodontic elastomeric chains.正畸弹性链的拉伸性能。
Eur J Orthod. 2004 Apr;26(2):157-62. doi: 10.1093/ejo/26.2.157.
9
The effect of ligation method on friction in sliding mechanics.结扎方法对滑动力学中摩擦力的影响。
Am J Orthod Dentofacial Orthop. 2003 Apr;123(4):416-22. doi: 10.1067/mod.2003.14.
10
Strength decay of orthodontic elastomeric ligatures.正畸弹性结扎丝的强度衰减
J Orthod. 2002 Mar;29(1):37-43. doi: 10.1093/ortho/29.1.37.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验