J Ultrasound Med. 2009 Nov;28(11):1511-8. doi: 10.7863/jum.2009.28.11.1511.
Volumetric flow measurement with Doppler ultrasound is useful in assessing blood flow as part of an evaluation of arteriovenous fistula maturity in patients undergoing hemodialysis. In this study, we assessed both accuracy and variability in volumetric flow measurements obtained using modern and commercially available ultrasound systems and an in vitro experimental setup.
Volumetric flow measurements using duplex ultrasound were obtained by 3 users operating 5 different systems for randomized flow in the range of 100 to 1000 mL/min. Users performed 3 consecutive measurements at a given flow rate. Data were analyzed using statistical techniques to assess measurement accuracy and variability.
Over the span of flow rates studied, the root mean square error (RMSE) for the 5 ultrasound systems ranged from 38.8 to 79.7, 36.8 to 52.0, 73.0 to 85.3, 26.7 to 44.6, and 43.9 to 93.5 mL/min. Corresponding average RMSE values were 60.3, 42.7, 81.1, 37.2, and 64.4 mL/min, respectively. A linear regression analysis of mean interobserver measurements revealed an excellent correlation for all ultrasound systems (r(2) > 99.1%). Assessment of intraobserver measurements revealed no statistically significant differences for any ultrasound system evaluated (P > .94). Comparison of interobserver measurements indicates no statistically significant differences between any of the 5 systems (P > .14).
Modern ultrasound systems are reasonably accurate in blood flow measurement in an experimental setup mimicking clinically relevant blood flow ranges in a hemodialysis fistula. Users need adequate training and experience to perform multiple measurements and use appropriate techniques to minimize errors in flow measurement.
多普勒超声容积流量测量可用于评估血流,作为评估接受血液透析患者动静脉瘘成熟度的一部分。本研究评估了现代和市售超声系统以及体外实验装置在容积流量测量中的准确性和可变性。
3 名使用者使用 5 种不同的系统对 100 至 1000 毫升/分钟范围内的随机流量进行双工超声容积流量测量。使用者在给定的流速下进行 3 次连续测量。使用统计技术分析数据,以评估测量的准确性和可变性。
在所研究的流速范围内,5 种超声系统的均方根误差(RMSE)范围为 38.8 至 79.7、36.8 至 52.0、73.0 至 85.3、26.7 至 44.6 和 43.9 至 93.5 毫升/分钟。相应的平均 RMSE 值分别为 60.3、42.7、81.1、37.2 和 64.4 毫升/分钟。所有超声系统的平均观察者间测量的线性回归分析均显示出极好的相关性(r(2)>99.1%)。观察者内测量的评估表明,评估的任何超声系统均无统计学上显著的差异(P>.94)。观察者间测量的比较表明,5 种系统之间无统计学上显著的差异(P>.14)。
在模拟临床相关血液透析瘘中血流范围的实验装置中,现代超声系统在血流测量方面具有相当的准确性。使用者需要进行足够的培训和经验,以进行多次测量,并使用适当的技术来最小化流量测量中的误差。