Department of Radiology, Peking University People's Hospital, Beijing, China.
Korean J Radiol. 2009 Nov-Dec;10(6):568-74. doi: 10.3348/kjr.2009.10.6.568.
To evaluate a quantitative method for individually adjusting the tube current to obtain images with consistent noise in electrocardiogram (ECG)-gated CT cardiac scans.
The image noise from timing bolus and cardiac CT scans of 80 patients (Group A) who underwent a 64-row multidetector (MD) CT cardiac examination with patient-independent scan parameters were analyzed. A formula was established using the noise correlation between the timing bolus and cardiac scans. This formula was used to predict the required tube current to obtain the desired cardiac CT image noise based on the timing bolus noise measurement. Subsequently, 80 additional cardiac patients (Group B) were scanned with individually adjusted tube currents using an established formula to evaluate its ability to obtain accurate and consistent image noise across the patient population. Image quality was evaluated using score scale of 1 to 5 with a score of 3 or higher being clinically acceptable.
Using the formula, we obtained an average CT image noise of 28.55 Hounsfield unit (HU), with a standard deviation of only 1.7 HU, as opposed to a target value of 28 HU. Image quality scores were 4.03 and 4.27 for images in Groups A and B, respectively, and there was no statistical difference between the image quality scores between the two groups. However, the average CT dose index (CTDIvol) was 30% lower for Group B.
Adjusting the tube current based on timing bolus scans may provide a consistent image quality and dose optimization for cardiac patients of various body mass index values.
评估一种定量方法,以便根据个体调整管电流,从而在心电图(ECG)门控 CT 心脏扫描中获得具有一致噪声的图像。
对 80 例行 64 排多层 CT(MD)心脏检查的患者(A 组)的定时团注和心脏 CT 扫描的图像噪声进行分析,这些患者的扫描参数不依赖于患者个体。利用定时团注和心脏扫描之间的噪声相关性建立了一个公式,该公式用于预测获得期望心脏 CT 图像噪声所需的管电流,基于定时团注噪声测量值。随后,使用建立的公式对 80 例额外的心脏患者(B 组)进行个体调整管电流扫描,以评估其在整个患者群体中获得准确且一致的图像噪声的能力。使用 1 到 5 的评分量表评估图像质量,评分 3 或更高为临床可接受。
使用该公式,我们获得了平均 CT 图像噪声为 28.55 亨氏单位(HU),标准偏差仅为 1.7 HU,而目标值为 28 HU。A 组和 B 组的图像质量评分分别为 4.03 和 4.27,两组之间的图像质量评分无统计学差异。然而,B 组的平均 CT 剂量指数(CTDIvol)降低了 30%。
基于定时团注扫描调整管电流可以为各种体重指数值的心脏患者提供一致的图像质量和剂量优化。