Int J Cardiovasc Imaging. 2011 Apr;27(4):611-7. doi: 10.1007/s10554-010-9701-5. Epub 2010 Sep 18.
To evaluate the clinical value of a body mass index (BMI) based tube current (mA) selection method for obtaining consistent image quality with dose optimization in MDCT prospective ECG gated coronary calcium scoring. A formula for selecting mA to achieve desired image quality based on patient BMI was established using a control group (A) of 200 MDCT cardiac patients with a standard scan protocol. One hundred patients in Group B were scanned with this BMI-dependent mA for achieving a desired noise level of 18 HU at 2.5 mm slice thickness. The CTDIvol and image noise on the ascending aorta for the two groups were recorded. Two experienced radiologists quantitatively evaluated the image quality using scores of 1-4 with 4 being the highest. The image quality scores had no statistical difference (P = 0.71) at 3.89 ± 0.32, 3.87 ± 0.34, respectively, for groups A and B of similar BMI. The image noise in Group A had linear relationship with BMI. The image noise in Group B using BMI-dependent mA was independent of BMI with average value of 17.9 HU and smaller deviations for the noise values than in Group A (2.0 vs. 2.9 HU). There was a 35% dose reduction with BMI-dependent mA selection method on average with the lowest effective dose being only 0.35 mSv for patient with BMI of 18.3. A quantitative BMI-based mA selection method in MDCT prospective ECG gated coronary calcium scoring has been proposed to obtain a desired and consistent image quality and provide dose optimization across patient population.
为了评估一种基于体重指数(BMI)的管电流(mA)选择方法在 MDCT 前瞻性 ECG 门控冠状动脉钙评分中实现剂量优化的一致性图像质量的临床价值。使用标准扫描方案对 200 例 MDCT 心脏患者的对照组(A 组)建立了一种基于患者 BMI 选择 mA 以达到预期图像质量的公式。B 组的 100 例患者使用这种基于 BMI 的 mA 以达到 2.5mm 切片厚度时 18HU 的预期噪声水平。记录两组的 CTDIvol 和升主动脉图像噪声。两名有经验的放射科医生使用 1-4 分的评分对图像质量进行定量评估,4 分为最高分。两组的图像质量评分无统计学差异(P=0.71),分别为 3.89±0.32 和 3.87±0.34。A 组的图像噪声与 BMI 呈线性关系。B 组使用基于 BMI 的 mA 的图像噪声与 BMI 无关,平均为 17.9HU,噪声值的偏差小于 A 组(2.0 比 2.9 HU)。使用基于 BMI 的 mA 选择方法,平均剂量减少了 35%,BMI 为 18.3 的患者的最低有效剂量仅为 0.35mSv。已经提出了一种 MDCT 前瞻性 ECG 门控冠状动脉钙评分的定量 BMI 基 mA 选择方法,以获得预期和一致的图像质量,并为患者群体提供剂量优化。