青光眼镜视神经头地形测量的可重复性。

Reproducibility of optic nerve head topographic measurements with the glaucoma-scope.

机构信息

Joseph M. Bryan Glaucoma Clinic, Duke University Eye Center, Durham, North Carolina, U.S.A.

出版信息

J Glaucoma. 1995 Jun;4(3):170-6.

Abstract

PURPOSE

Our purpose was to evaluate the reproducibility of optic nerve head depth measurements using the Glaucoma-Scope.

MATERIALS AND METHODS

Optic nerve head topography was measured with the Glaucoma-Scope in 10 normal subjects and in 10 patients with glaucoma. Five separate readings were obtained by a single observer from one eye of each subject on the same day. Mean depth values were calculated for the overall disc, each quadrant, and areas overlying vessels at those points at which the change from baseline exceeded 50 mum. Similar calculations were made for actual depth measurements in regions of potential clinical interest, including the bottom of the cup, the temporal slope, the nasal disc, and a peripapillary region, as well as 15 randomly selected single points. Reproducibility of the depth values was determined by calculating standard deviations, coefficients of variation, and 95% confidence intervals in the five separate measurements.

RESULTS

The mean standard deviations in regions of potential clinical interest among the 20 subjects ranged from 3.98 mum to 14.09 mum, and the mean coefficient of variation ranged between 4.2% and 8.7%. The 95% confidence intervals for the total population ranged from 9.88 mum to 34.97 mum for the regions of interest and was 43 mum for the 15 random points. There was no statistically significant difference in reproducibility between the two groups of subjects.

CONCLUSION

In this study the Glaucoma-Scope produced quantitative optic nerve head topographic measurements that were highly reproducible, within a tolerance of <50 mum.

摘要

目的

我们的目的是评估使用 Glaucoma-Scope 进行视神经头深度测量的可重复性。

材料和方法

使用 Glaucoma-Scope 对 10 名正常受试者和 10 名青光眼患者的视神经头进行了测量。同一位观察者在同一天对每位受试者的一只眼进行了 5 次单独测量。为整个盘、每个象限以及在基线变化超过 50μm 的那些点上覆盖血管的区域计算了平均深度值。在潜在临床感兴趣的区域(包括杯底、颞侧斜坡、鼻盘和视盘周围区域)以及 15 个随机选择的单点,也进行了类似的实际深度测量计算。通过计算 5 次单独测量的标准偏差、变异系数和 95%置信区间,确定深度值的可重复性。

结果

在 20 名受试者中,潜在临床感兴趣区域的平均标准偏差范围为 3.98μm 至 14.09μm,平均变异系数范围为 4.2%至 8.7%。总人群的 95%置信区间为 9.88μm 至 34.97μm,用于感兴趣区域,15 个随机点为 43μm。两组受试者之间的可重复性无统计学差异。

结论

在这项研究中,Glaucoma-Scope 产生了高度可重复的定量视神经头地形图测量值,在<50μm 的容差内。