组织血红蛋白指数:一种非侵入性的组织总血红蛋白光学测量方法。

Tissue hemoglobin index: a non-invasive optical measure of total tissue hemoglobin.

机构信息

BioMeasurement Division, Hutchinson Technology Inc, 40 West Highland Park Drive NE, Hutchinson, MN 55350-9784, USA.

出版信息

Crit Care. 2009;13 Suppl 5(Suppl 5):S2. doi: 10.1186/cc8000. Epub 2009 Nov 30.

Abstract

INTRODUCTION

The tissue hemoglobin index (THI) is a hemoglobin signal strength metric provided on the InSpectra StO2 Tissue Oxygenation Monitor, Model 650. There is growing interest regarding the physiologic meaning of THI and whether a clinically useful correlation between THI and blood hemoglobin concentration exists. A series of in vitro and in vivo experiments was performed to evaluate whether THI has potential utility beyond its primary purpose of helping InSpectra device users optimally position a StO2 sensor over muscle tissue.

METHODS

The THI and tissue hemoglobin oxygen saturation (StO2) were measured using the InSpectra StO2 Tissue Oxygenation Monitor, Model 650, with a 15 mm optical sensor. A THI normal reference range was established in the thenar eminence (hand) for 434 nonhospitalized human volunteers. In 30 subjects, the thenar THI was also evaluated during 5-minute arterial and venous blood flow occlusions, and with blood volume exsanguination in the hand induced with an Esmarch bandage. In addition, correlation of the THI to blood total hemoglobin concentration (Hbt) was studied in five pigs whose Hbt was isovolumetrically diluted from 13 to 4 g/dl systemically and 0.5 g/dl locally in the hind limb. The sensitivity and specificity of the THI to measure tissue hemoglobin concentration (THC) were characterized in vitro using isolated blood tissue phantoms.

RESULTS

In human thenar tissue, the average THI was 14.1 +/- 1.6 (mean +/- standard deviation). The THI extrapolated to 100% blood volume exsanguination was 3.7 +/- 2.0 units presumably from myoglobin. On average, the THI increased 1.5 +/- 1.0 units with venous occlusion and decreased 4.0 +/- 2.0 units with arterial occlusion. In porcine hind limbs, the THI weakly correlated with Hbt (r2 = 0.26) while DeltaTHI during venous occlusion had a stronger correlation (r2 = 0.62). In vitro tests indicated that THI strongly correlated (r2 > 0.99) to phantom THC and was insensitive to StO2 changes.

CONCLUSIONS

Steady-state THI values do not reliably indicate Hbt. The THI is a reproducible quantitative index for THC, and THI trends can discriminate between arterial or venous blood flow occlusions. The THI magnitude permits the estimation of myoglobin's contribution to StO2.

摘要

简介

组织血红蛋白指数(THI)是 InSpectra StO2 组织氧饱和度监测仪模型 650 提供的血红蛋白信号强度指标。越来越多的人关注 THI 的生理意义,以及 THI 是否与血液血红蛋白浓度存在临床有用的相关性。进行了一系列的体外和体内实验,以评估 THI 是否除了帮助 InSpectra 设备用户优化 StO2 传感器在肌肉组织上的位置之外,是否具有潜在的用途。

方法

使用 InSpectra StO2 组织氧饱和度监测仪模型 650 和 15 毫米光学传感器测量 THI 和组织血红蛋白氧饱和度(StO2)。在 434 名非住院的人类志愿者的大鱼际(手)中建立了 THI 正常参考范围。在 30 名受试者中,还评估了大鱼际的 THI 在 5 分钟动脉和静脉血流闭塞期间以及通过 Esmarch 绷带在手诱导的血液体积放血期间的情况。此外,还研究了 THI 与 5 头猪的全血血红蛋白浓度(Hbt)的相关性,这些猪的 Hbt 通过系统地将 13 至 4 g/dl 系统和 0.5 g/dl 局部稀释到后肢来等容稀释。使用离体血液组织体模在体外对 THI 测量组织血红蛋白浓度(THC)的灵敏度和特异性进行了表征。

结果

在人类大鱼际组织中,平均 THI 为 14.1 +/- 1.6(平均值 +/- 标准差)。THI 外推至 100%血液体积放血时为 3.7 +/- 2.0 单位,可能来自肌红蛋白。平均而言,静脉闭塞时 THI 增加 1.5 +/- 1.0 单位,动脉闭塞时减少 4.0 +/- 2.0 单位。在猪的后肢中,THI 与 Hbt 弱相关(r2 = 0.26),而静脉闭塞期间的 DeltaTHI 相关性更强(r2 = 0.62)。体外测试表明,THI 与体模 THC 高度相关(r2 > 0.99),并且对 StO2 变化不敏感。

结论

稳态 THI 值不能可靠地指示 Hbt。THI 是 THC 的可重复定量指标,THI 趋势可区分动脉或静脉血流闭塞。THI 幅度允许估计肌红蛋白对 StO2 的贡献。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/00bb/2786104/ae094c1676af/cc8000-1.jpg