J Occup Environ Hyg. 2010 Feb;7(2):88-93. doi: 10.1080/15459620903418746.
The Institute of Occupational Medicine (IOM) personal sampler is usually operated at a flow rate of 2.0 L/min, the rate at which it was designed and calibrated, for sampling the inhalable mass fraction of airborne particles in occupational environments. In an environment of low aerosol concentrations only small amounts of material are collected, and that may not be sufficient for analysis. Recently, a new sampling pump with a flow rate up to 15 L/min became available for personal samplers, with the potential of operating at higher flow rates. The flow rate of a Leland Legacy sampling pump, which operates at high flow rates, was evaluated and calibrated, and its maximum flow was found to be 10.6 L/min. IOM samplers were placed on a mannequin, and sampling was conducted in a large aerosol wind tunnel at wind speeds of 0.56 and 2.22 m/s. Monodisperse aerosols of oleic acid tagged with sodium fluorescein in the size range of 2 to 100 microm were used in the test. The IOM samplers were operated at flow rates of 2.0 and 10.6 L/min. Results showed that the IOM samplers mounted in the front of the mannequin had a higher sampling efficiency than those mounted at the side and back, regardless of the wind speed and flow rate. For the wind speed of 0.56 m/s, the direction-averaged (the average value of all orientations facing the wind direction) sampling efficiency of the samplers operated at 2.0 L/min was slightly higher than that of 10.6 L/min. For the wind speed of 2.22 m/s, the sampling efficiencies at both flow rates were similar for particles < 60 microm. The results also show that the IOM's sampling efficiency at these two different flow rates follows the inhalable mass curve for particles in the size range of 2 to 20 microm. The test results indicate that the IOM sampler can be used at higher flow rates.
职业医学研究所(IOM)个人采样器通常以 2.0 L/min 的流速运行,这是其设计和校准的流速,用于采样职业环境中空气传播颗粒的可吸入质量分数。在气溶胶浓度低的环境中,只有少量的物质被收集,这可能不足以进行分析。最近,一种新的采样泵,流速可达 15 L/min,可用于个人采样器,并有可能以更高的流速运行。对一种可高速运行的 Leland Legacy 采样泵的流速进行了评估和校准,发现其最大流速为 10.6 L/min。将 IOM 采样器放置在人体模型上,并在风速为 0.56 和 2.22 m/s 的大型气溶胶风洞中进行采样。测试中使用了带有荧光素钠标记的油酸单分散气溶胶,粒径范围为 2 至 100 微米。IOM 采样器以 2.0 和 10.6 L/min 的流速运行。结果表明,无论风速和流速如何,安装在人体模型前部的 IOM 采样器的采样效率均高于安装在侧面和背面的采样器。对于风速为 0.56 m/s,以 2.0 L/min 运行的采样器的方向平均(面向风向的所有方向的平均值)采样效率略高于 10.6 L/min。对于风速为 2.22 m/s,两种流速下的采样效率对于 <60 微米的颗粒相似。结果还表明,IOM 在这两种不同流速下的采样效率遵循 2 至 20 微米粒径范围内颗粒的可吸入质量曲线。测试结果表明,IOM 采样器可以在更高的流速下使用。