双眼植入+3.0D 附加度数消像散衍射型非球面多焦点人工晶状体后的视觉功能。

Visual function after bilateral implantation of apodized diffractive aspheric multifocal intraocular lenses with a +3.0 D addition.

机构信息

Department of Ophthalmology, Goethe-University, Frankfurt am Main, Germany.

出版信息

J Cataract Refract Surg. 2009 Dec;35(12):2062-9. doi: 10.1016/j.jcrs.2009.08.013.

Abstract

PURPOSE

To evaluate visual function after bilateral implantation of apodized diffractive aspheric multifocal intraocular lenses (IOLs) with a +3.0 diopter addition (add) power.

SETTING

Multicenter study at 5 European sites.

METHODS

Five surgeons prospectively enrolled patients to receive bilateral implantation of AcrySof IQ ReSTOR SN6AD1 IOLs. Assessments included defocus testing, uncorrected and corrected distance visual acuities at various distances, and patient questionnaires.

RESULTS

Ninety-three patients were enrolled. The mean distance-corrected visual acuities at far, intermediate, and near distances were significantly better postoperatively. At 6 months, uncorrected visual acuity (logMAR) was -0.03 +/- 0.13 (SD) at 4 m, 0.20 +/- 0.14 at 70 cm, 0.13 +/- 0.15 at 60 cm, 0.05 +/- 0.18 at 50 cm, and 0.04 +/- 0.11 at 40 cm. The mean patient-preferred near distance was 41 +/- 4 cm, at which distance the mean visual acuity was -0.01 +/- 0.11 logMAR. The defocus curve had a plateau of optimum near vision from 40 to 50 cm. Postoperatively, patients reported having minimal to no difficulty with 22 of 27 visual disturbances or visual activities; the other 5 items were ranked minimally to moderately difficult. The mean patient satisfaction with vision was 8.3 +/- 1.6 (out of 10); 88% of patients were spectacle independent.

CONCLUSIONS

Bilateral apodized diffractive aspheric multifocal IOLs with a +3.0 D add provided a broad range of optimum near vision, good intermediate visual acuity, and low rates of visual disturbances. Patients were highly satisfied with their vision, and 88% were spectacle independent.

摘要

目的

评估双眼植入+3.0 屈光度附加(add)的消色差衍射多焦点人工晶状体(IOL)后的视觉功能。

设置

欧洲 5 个地点的多中心研究。

方法

5 位外科医生前瞻性地招募患者接受 AcrySof IQ ReSTOR SN6AD1 IOL 的双眼植入。评估包括离焦测试、不同距离的未矫正和矫正远视力以及患者问卷调查。

结果

93 名患者入组。远、中、近距的平均距离矫正视力术后显著提高。6 个月时,未矫正视力(logMAR)在 4 m 处为-0.03 +/- 0.13(SD),在 70 cm 处为 0.20 +/- 0.14,在 60 cm 处为 0.13 +/- 0.15,在 50 cm 处为 0.05 +/- 0.18,在 40 cm 处为 0.04 +/- 0.11。患者首选的平均近距为 41 +/- 4 cm,在此距离的平均视力为-0.01 +/- 0.11 logMAR。离焦曲线在 40 至 50 cm 之间有一个近视力最佳的平台。术后,患者报告 27 种视觉障碍或视觉活动中有 22 种只有轻微至无困难;其他 5 项被评为轻微至中度困难。患者对视力的平均满意度为 8.3 +/- 1.6(满分 10 分);88%的患者无需戴眼镜。

结论

双眼植入+3.0 D 的消色差衍射多焦点 IOL 提供了广泛的最佳近视力、良好的中间视力和较低的视觉障碍发生率。患者对他们的视力非常满意,88%的患者无需戴眼镜。