Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN 55905, USA.
Clin Chem. 2010 Feb;56(2):306-13. doi: 10.1373/clinchem.2009.134643. Epub 2009 Dec 10.
Immunoassays specific for 1-84 parathyroid hormone (PTH) reportedly reflect the bioactivity of PTH; however, PTH immunoassays can be susceptible to interference by cross-reacting PTH fragments. In addition, these assays currently lack standardization. A methodology using immunocapture purification with liquid chromatography-tandem mass spectrometry (LC-MS/MS) detection, along with a stable isotope-labeled internal standard, may help address these issues.
We isolated 1-84 PTH from 1 mL serum by immunocapture on a 6.5-mm polystyrene bead. The immobilized PTH was digested in situ and analyzed by LC-MS/MS. For quantification, we used the selected reaction monitoring response from the N-terminal tryptic peptide 1-13 PTH ((1)SVSEIQLMHNLGK(13)).
The linear range of the assay was 39.1-4560 ng/L, and the limit of detection and limit of quantification were 14.5 ng/L and 39.1 ng/L, respectively. The intraassay CVs ranged from 6% to 11%, and the interassay CVs ranged from 7% to 17%. Interference by PTH fragments 1-44 PTH, 7-84 PTH, 43-68 PTH, 52-84 PTH, 64-84 PTH, and PTH-related protein (PTHrP) was <or=1% to <or=0.001%. Method comparison of LC-MS/MS vs the Roche Cobas immunoassay yielded Deming fit of LC-MS/MS = 1.01x immunoassay - 13.21. The mean bias by Bland-Altman plot was -9.4%.
In patients with hyperparathyroidism, the immunocapture in situ digestion LC-MS/MS method can provide accurate and precise PTH results compared with immunoassay.
据称,针对甲状旁腺激素(PTH)1-84 位的免疫测定法能够反映 PTH 的生物活性;然而,这些免疫测定法可能容易受到交叉反应性 PTH 片段的干扰。此外,这些测定法目前缺乏标准化。一种使用免疫捕获纯化结合液相色谱-串联质谱(LC-MS/MS)检测,以及稳定同位素标记的内标方法,可能有助于解决这些问题。
我们通过在 6.5mm 聚苯乙烯珠上的免疫捕获,从 1mL 血清中分离出 1-84 PTH。将固定化的 PTH 在原位进行消化,并通过 LC-MS/MS 进行分析。为了定量,我们使用 N 端胰蛋白酶肽 1-13 PTH 的选择反应监测响应((1)SVSEIQLMHNLGK(13))。
该测定法的线性范围为 39.1-4560ng/L,检测限和定量限分别为 14.5ng/L 和 39.1ng/L。批内 CV 范围为 6%-11%,批间 CV 范围为 7%-17%。PTH 片段 1-44 PTH、7-84 PTH、43-68 PTH、52-84 PTH、64-84 PTH 和甲状旁腺激素相关蛋白(PTHrP)的干扰均<or=1%至<or=0.001%。LC-MS/MS 与罗氏 Cobas 免疫测定法的方法比较得出,LC-MS/MS 的 Deming 拟合值为 1.01x 免疫测定法-13.21。Bland-Altman 图的平均偏差为-9.4%。
在甲状旁腺功能亢进症患者中,与免疫测定法相比,原位消化 LC-MS/MS 免疫捕获法能够提供准确、精密的 PTH 结果。