采用 2D 电离室矩阵 i'mRT MatriXX 进行调强放射治疗(IMRT)的混合计划验证——一项可行性研究。

Hybrid plan verification for intensity-modulated radiation therapy (IMRT) using the 2D ionization chamber array I'mRT MatriXX--a feasibility study.

机构信息

Department of Radiotherapy, Regensburg University Medical Center, Regensburg, Germany.

出版信息

Phys Med Biol. 2010 Jan 21;55(2):N39-55. doi: 10.1088/0031-9155/55/2/N02. Epub 2009 Dec 21.

Abstract

The 2D ionization chamber array I'mRT MatriXX (IBA, Schwarzenbruck, Germany) has been developed for absolute 2D dosimetry and verification of intensity-modulated radiation therapy (IMRT) for perpendicular beam incidence. The aim of this study is to evaluate the applicability of I'mRT MatriXX for oblique beam incidence and hybrid plan verification of IMRT with original gantry angles. For the assessment of angular dependence, open fields with gantry angles in steps of 10 degrees were calculated on a CT scan of I'mRT MatriXX. For hybrid plan verification, 17 clinical IMRT plans and one rotational plan were used. Calculations were performed with pencil beam (PB), collapsed cone (CC) and Monte Carlo (MC) methods, which had been previously validated. Measurements were conducted on an Elekta SynergyS linear accelerator. To assess the potential and limitations of the system, gamma evaluation was performed with different dose tolerances and distances to agreement. Hybrid plan verification passed the gamma test with 4% dose tolerance and 3 mm distance to agreement in all cases, in 82-88% of the cases for tolerances of 3%/3 mm, and in 59-76% of the cases if 3%/2 mm were used. Separate evaluation of the low dose and high dose regions showed that I'mRT MatriXX can be used for hybrid plan verification of IMRT plans within 3% dose tolerance and 3 mm distance to agreement with a relaxed dose tolerance of 4% in the low dose region outside the multileaf collimator (MLC).

摘要

2D 电离室矩阵 I'mRT MatriXX(IBA,德国施瓦岑布鲁克)专为绝对 2D 剂量测定和强度调制放射治疗(IMRT)的垂直射束入射验证而开发。本研究的目的是评估 I'mRT MatriXX 用于斜射束入射和带有原始旋转机架角度的混合计划验证的适用性。为了评估角度依赖性,在 I'mRT MatriXX 的 CT 扫描上计算了具有旋转机架角度步长为 10 度的开放野。对于混合计划验证,使用了 17 个临床 IMRT 计划和 1 个旋转计划。计算使用了先前经过验证的笔形束(PB)、崩溃锥体(CC)和蒙特卡罗(MC)方法。测量是在 Elekta SynergyS 线性加速器上进行的。为了评估系统的潜力和局限性,使用不同的剂量容限和符合距离进行了伽马评估。在所有情况下,混合计划验证都通过了伽马测试,使用 4%的剂量容限和 3 毫米的符合距离,对于 3%/3 毫米的容限,通过率为 82-88%,对于 3%/2 毫米的容限,通过率为 59-76%。对低剂量区和高剂量区的单独评估表明,I'mRT MatriXX 可以用于混合计划验证,对于 IMRT 计划,在低剂量区(多叶准直器(MLC)外)使用 4%的宽松剂量容限,在 3%的剂量容限和 3 毫米的符合距离内,都可以通过验证。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验