AJR Am J Roentgenol. 2010 Jan;194(1):W115-8. doi: 10.2214/AJR.09.2927.
We show with both a clinical case and measurements using the American College of Radiology mammography phantom that some digital mammography acquisition and image display combinations lead to both marked over- and underestimation of lesion sizes on geometrically magnified images.
The results of this study indicate that the accuracy of lesion size measurements in all magnification modes should be a routine quality control acceptance test with each acquisition-display system combination in digital mammography.