Suppr超能文献

含 K2EDTA 抗凝剂的未充满血液采集管可用于自动化全血细胞计数、白细胞分类计数和网织红细胞计数。

Under-filled blood collection tubes containing K2EDTA as anticoagulant are acceptable for automated complete blood counts, white blood cell differential, and reticulocyte count.

机构信息

Department of Laboratories, Seattle Children's Hospital, Seattle, WA 98105, USA.

出版信息

Int J Lab Hematol. 2010 Oct;32(5):491-7. doi: 10.1111/j.1751-553X.2009.01211.x. Epub 2009 Dec 23.

Abstract

Current laboratory standards from Clinical Laboratory Standards Institute (CLSI) and manufacturer's (Becton Dickinson) data indicate that under-filling K(2)EDTA blood collection tubes can result in erroneous hematology values. To accommodate under-filled tubes and reduce collection volumes while optimizing our automation, we explored the acceptable limit of under-filled tubes for hematology values. We collected 8.0 ml of blood from 30 normal adult volunteers. Each donation was aliquoted in the following volumes: 4.0, 2.0, 1.0, 0.5 ml x 2. These samples were analyzed within 1 h of blood collection on Sysmex XE-2100 (Sysmex America Inc., Mundelein, IL, USA) for complete blood count, reticulocyte, and white blood cell differentials. Results of the under-filled tubes were compared to those of the standard volume. The Deming regression analysis show excellent correlation for all parameters between each under-filled blood collection volume compared to a standard 4 ml volume. The Bland and Altman analysis shows good agreement between both 1.0 and 2.0 ml compared to a 4.0 ml volume. The 0.5 ml compared to a 4.0 ml volume, however, shows increased variation on many parameters. In addition all three collection volumes show negative bias compared to the standard volume for platelet count, but the difference is considered insignificant with a percent difference of 5.5%, 3.2%, and 1.5% for 0.5, 1.0, and 2.0 ml collection volume respectively. Finally for 0.5 ml collection volume we noticed a low level of false positive flagging rate for white blood cell. Acceptable complete blood count values of under-filled powdered K(2)EDTA tubes can be obtained with as little as 1.0 ml of blood.

摘要

目前,临床实验室标准协会(CLSI)和制造商(BD)的数据表明,K2EDTA 血液采集管未充满会导致血液学值出现错误。为了适应未充满的采血管并在优化自动化操作的同时减少采集量,我们探索了血液学值可接受的未充满采血管下限。我们从 30 名正常成年志愿者中采集了 8.0 毫升血液。每个献血者的样本均按以下体积等分:4.0、2.0、1.0、0.5 ml×2。这些样本在采血后 1 小时内使用 Sysmex XE-2100(Sysmex America Inc.,Mundelein,IL,USA)进行全血细胞计数、网织红细胞和白细胞分类分析。将未充满的采血管结果与标准体积的结果进行比较。Deming 回归分析显示,与标准 4 ml 体积相比,所有参数在每个未充满的采血管体积之间均具有极好的相关性。Bland 和 Altman 分析显示,与 4.0 ml 体积相比,1.0 和 2.0 ml 体积之间具有良好的一致性。然而,与 4.0 ml 体积相比,0.5 ml 体积在许多参数上显示出更大的变化。此外,与标准体积相比,所有三个采集体积的血小板计数都存在负偏倚,但差异被认为不显著,分别为 0.5、1.0 和 2.0 ml 采集体积的 5.5%、3.2%和 1.5%。最后,对于 0.5 ml 采集体积,我们注意到白细胞的假阳性标记率较低。使用少至 1.0 毫升的血液即可获得可接受的未充满粉末状 K2EDTA 采血管的全血细胞计数值。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验