Department of Pediatric Retina, Narayana Nethralaya Postgraduate Institute of Ophthalmology, Bangalore, India.
Eye (Lond). 2010 Feb;24(2):379-82. doi: 10.1038/eye.2009.313. Epub 2010 Jan 8.
To acquire high definition spectral domain optical coherence tomography (SD-OCT) images of supine, non-anesthetized infants in the office by a novel modification of a commercially available table-top, chin rest system.
Spectralis, a combined HRA+OCT device (Heidelberg Engineering, Heidelberg, Germany) was modified to convert the table-top system into a hand-held device using a two-step modification of the existing system. This device was used to obtain high-resolution OCT images of four infants with aggressive posterior retinopathy of prematurity (APROP) who required no general anesthesia.
Serial imaging at two time periods of the exact area of clinical interest were obtained and were comparable. The entire procedure was safely completed in the office. The obtained OCT images guided selective laser ablation of flat neovascularization which were missed on clinical examination.
With this simple modification and technique, the ability of using the Spectralis, a popular combined angiography and OCT imaging device for imaging supine, non-anesthetized infants has been established. This possibility would allow the dual use of the table-top system to serve also as a hand-held device for paediatric cases that can be imaged in the office setting with limited operating room facility in a busy practice.
通过对商业台式、下巴托系统的新型改造,实现在诊室内获取仰卧、非麻醉婴儿的高清晰度谱域光学相干断层扫描(SD-OCT)图像。
使用两步法对 Spectralis(海德堡工程公司,德国海德堡)这款集高分辨率血管造影和 OCT 成像于一体的设备进行改造,将其转化为手持设备。使用该设备对 4 名患有严重早产儿后部视网膜病变(APROP)的婴儿进行高分辨率 OCT 成像,这些婴儿无需全身麻醉。
在两次不同时间点对同一临床感兴趣区域进行了连续成像,且结果具有可比性。整个过程在诊室内安全完成。获得的 OCT 图像引导了对平坦新生血管的选择性激光消融,这些新生血管在临床检查中被遗漏。
通过这种简单的改进和技术,成功实现了使用 Spectralis(一种流行的集血管造影和 OCT 成像于一体的设备)对仰卧、非麻醉婴儿进行成像的能力。这种可能性允许台式系统的双重用途,也可作为一种手持设备,用于在繁忙的诊室内,在有限的手术室条件下,对儿科病例进行成像。