Mass Spectrometry Facility and Department of Pharmaceutical Chemistry, University of California, San Francisco, San Francisco, California 94158-2517, USA.
J Am Soc Mass Spectrom. 2010 Mar;21(3):455-9. doi: 10.1016/j.jasms.2009.11.011. Epub 2009 Dec 11.
Stored waveform inverse Fourier transform (SWIFT) technology has been implemented on a commercial Fourier transform ICR mass spectrometer. Complex ejection/isolation waveforms can be generated on an arbitrary waveform generator (AWG) and applied to the instrument by use of a high speed analog switch. High mass selectivity and subsequent electron capture dissociation (ECD) of the SWIFT isolated ions has been demonstrated with analysis of intact Bovine histone H4. A mass selectivity about 0.1 m/z unit for isolation of the 18+ charge state ions was achieved. Adaptation of SWIFT on the commercial instrument provides an enhanced capability for characterizing intact proteins by top-down analysis.