Vasconcelos de Moraes Carlos Gustavo, Castro Reis Alexandre Soares, Sano Milena Eimi, Barreira Alan Kardec, Vessani Roberto Murad, Jr Susanna Remo
J Ocul Biol Dis Infor. 2009 Feb 24;2(1):29-32. doi: 10.1007/s12177-009-9016-8.
The aim of this study was to compare the intraocular pressure (IOP) profile during the modified diurnal tension curve (mDTC) using Goldman applanation tonometry (GAT) and dynamic contour tonometry (DCT) in treated glaucomatous eyes. Eligible subjects were submitted to the mDTC using GAT and DCT in this sequence. IOP measurements were performed at 8 a.m., 10 a.m., 2 p.m., and 4 p.m.. Central corneal thickness was measured using ultrasound pachymetry in the morning. Statistical analysis was performed using paired Student's t test and Bland-Altman plot. The mean difference between DCT and GAT measurements was 0.9 mmHg. The mean +/- SD IOP measurements during the mDTC were 19.68 +/- 4.68, 17.63 +/- 4.44, 17.25 +/- 5.41, and 17.32 +/- 4.25 mmHg using GAT and 19.97 +/- 4.75, 18.79 +/- 4.61, 19.53 +/- 5.30, and 19.43 +/- 5.45 mmHg using DCT. IOP measurements were higher in the morning (8 a.m.) and decreased throughout the day using both tonometers. The difference between IOP measurements using GAT and DCT was smaller in the morning and increased throughout the day. The IOP variability using GAT was higher than using DCT. Corneal biomechanical properties might help explain our findings.
本研究的目的是比较在接受治疗的青光眼眼中,使用戈德曼压平眼压计(GAT)和动态轮廓眼压计(DCT)进行改良昼夜眼压曲线(mDTC)检查期间的眼压(IOP)情况。符合条件的受试者按此顺序使用GAT和DCT进行mDTC检查。眼压测量在上午8点、10点、下午2点和4点进行。上午使用超声测厚仪测量中央角膜厚度。使用配对学生t检验和布兰德-奥特曼图进行统计分析。DCT和GAT测量值之间的平均差值为0.9 mmHg。使用GAT时,mDTC期间的平均±标准差眼压测量值分别为19.68±4.68、17.63±4.44、17.25±5.41和17.32±4.25 mmHg,使用DCT时分别为19.97±4.75、18.79±4.61、19.53±5.30和19.43±5.45 mmHg。两种眼压计测量的眼压在上午(8点)均较高,并在全天逐渐下降。使用GAT和DCT测量的眼压差值在上午较小,全天逐渐增大。使用GAT时的眼压变异性高于使用DCT时。角膜生物力学特性可能有助于解释我们的研究结果。