Suppr超能文献

使用楔形固体模体结合射野轮廓扫描仪对电子束进行能量恒定性检查。

Energy constancy checking for electron beams using a wedge-shaped solid phantom combined with a beam profile scanner.

作者信息

Rosenow U F, Islam M K, Gaballa H, Rashid H

机构信息

Department of Clinical Radiobiology and Clinical Radiation Physics, University of Göttingen, Federal Republic of Germany.

出版信息

Med Phys. 1991 Jan-Feb;18(1):19-25. doi: 10.1118/1.596719.

Abstract

An energy constancy checking method is presented which involves a specially designed wedge-shaped solid phantom in combination with a multiple channel ionization chamber array known as the Thebes device. Once the phantom/beam scanner combination is set up, measurements for all electron energies can be made and evaluated without re-entering the treatment room. This is also valid for the readjustment of beam energies which are found to deviate from required settings. The immediate presentation of the measurements is in the form of crossplots which resemble depth dose profiles. The evaluation of the measured data can be performed using a hand-held calculator, but processing of the measured signals through a PC-type computer is advisable. The method is insensitive to usual fluctuations in beam flatness. The sensitivity and reproducibility of the method are more than adequate. The method may also be used in modified form for photon beams.

摘要

本文介绍了一种能量恒定检查方法,该方法涉及一个专门设计的楔形固体模体,与一种称为底比斯装置的多通道电离室阵列相结合。一旦设置好模体/束扫描器组合,就可以对所有电子能量进行测量和评估,而无需重新进入治疗室。对于发现偏离所需设置的束能量的重新调整,这同样有效。测量结果立即以类似于深度剂量分布的交叉图形式呈现。测量数据的评估可以使用手持计算器进行,但建议通过个人计算机类型的计算机对测量信号进行处理。该方法对束平整度的常见波动不敏感。该方法的灵敏度和再现性绰绰有余。该方法也可以以修改后的形式用于光子束。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验