Suppr超能文献

单体素 MRS 采用前瞻性运动校正和回顾性频率校正。

Single-voxel MRS with prospective motion correction and retrospective frequency correction.

机构信息

Medical Physics, Department of Diagnostic Radiology, University Hospital of Freiburg, Freiburg, Germany.

出版信息

NMR Biomed. 2010 Apr;23(3):325-32. doi: 10.1002/nbm.1469.

Abstract

Subject motion during MRS investigations is a factor limiting the quality and the diagnostic value of the spectra. The possibility of using external motion tracking data to correct for artefacts in MR imaging has been demonstrated previously. In this paper the utility of prospective motion correction for single-voxel proton MRS is investigated. The object motion data are used in real time to update the position of the spectroscopy voxel during the acquisition prior to every sequence repetition cycle. It is not, however, sufficient to update the voxel position alone due to shim changes accompanying subject motion. Adverse effects of frequency shifts induced by subject motion are effectively suppressed by the interleaved reference scan method.

摘要

在 MRS 研究中,受试者运动是限制谱质量和诊断价值的一个因素。先前已经证明,可以使用外部运动跟踪数据来校正磁共振成像中的伪影。在本文中,研究了单体素质子 MRS 中前瞻性运动校正的实用性。在每个序列重复周期之前,在采集期间,将物体运动数据实时用于更新光谱体素的位置。然而,由于受试者运动伴随的匀场变化,仅更新体素位置是不够的。通过交错参考扫描方法有效地抑制了由受试者运动引起的频率偏移的不利影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验