ABX advanced biochemical compounds, Radeberg, Germany.
Phys Med Biol. 2010 Feb 21;55(4):1099-113. doi: 10.1088/0031-9155/55/4/013. Epub 2010 Jan 28.
We studied quantitatively the effects of the discontinuity introduced in an otherwise homogeneous background by the cold walls of the standard spherical glass inserts commonly used in phantom measurements for calibration of threshold-based approaches to volumetric evaluation of PET investigations. We concentrated especially on the question of threshold-based volume determination. We computed analytically the convolution of an isotropic Gaussian point-spread function with the insert geometry (hot sphere + cold wall + warm background) and derived the theoretical background dependence of the volume reproducing threshold. This analysis shows a clear wall-related reduction of the optimal threshold with increasing background. The predictions of our theoretical analysis were verified in phantom measurements at background fractions between 0 and 0.29. Defining the background-corrected relative threshold [formula: see text] (T(abs): absolute volume reproducing threshold, A: measured activity at centre, B: background), we find that for a wall-less sphere T is independent of the background level. In the presence of cold walls, T drops (for not too small spheres, where recovery at the centre approaches 100%) from about 43% at B/A = 0 to about 25% at B/A = 0.5. Applying these thresholds to wall-less spheres leads to sizeable overestimates of the true volumes (43% at B/A = 0.5 for a sphere of 6 ml volume). We conclude that phantom measurements with standard sphere inserts for calibration of optimal thresholding algorithms introduce a systematic bias if performed at finite background levels. The observed background dependence is an artefact of the measurement procedure and does not reflect the conditions present in actual patient investigations.
我们定量研究了标准球形玻璃插入物冷壁在同质背景中引入的不连续性对基于阈值的体素评估方法校准 PET 研究体积的影响。我们特别关注基于阈值的体积确定问题。我们对各向同性高斯点扩散函数与插入物几何形状(热球+冷壁+暖背景)的卷积进行了分析,并推导出了体积再现阈值的理论背景依赖性。该分析表明,随着背景的增加,最佳阈值与壁相关的降低是显而易见的。我们的理论分析预测在背景分数为 0 到 0.29 之间的体模测量中得到了验证。定义背景校正的相对阈值 [公式:见文本](T(abs):绝对体积再现阈值,A:中心处的测量活性,B:背景),我们发现对于无壁球体,T 不依赖于背景水平。在存在冷壁的情况下,T 从 B/A = 0 时的约 43%下降到 B/A = 0.5 时的约 25%(对于中心处的回收率不太小的球体,接近 100%)。将这些阈值应用于无壁球体将导致对真实体积的估计过大(对于 6ml 体积的球体,B/A = 0.5 时为 43%)。我们得出结论,如果在有限的背景水平下进行测量,则使用标准球体插入物进行最佳阈值算法校准的体模测量会引入系统偏差。观察到的背景依赖性是测量过程的一种人为产物,并不反映实际患者研究中的条件。