使用数字式Biograph Vision PET/CT对F、Ga和Cu体模测量的图像质量和空间分辨率进行比较。

Comparison of image quality and spatial resolution between F, Ga, and Cu phantom measurements using a digital Biograph Vision PET/CT.

作者信息

Braune Anja, Oehme Liane, Freudenberg Robert, Hofheinz Frank, van den Hoff Jörg, Kotzerke Jörg, Hoberück Sebastian

机构信息

Department of Nuclear Medicine, University Hospital Carl Gustav Carus at the Technische Universität Dresden, Dresden, Germany.

PET Center, Institute of Radiopharmaceutical Cancer Research, Helmholtz-Zentrum Dresden-Rossendorf, Dresden, Germany.

出版信息

EJNMMI Phys. 2022 Sep 5;9(1):58. doi: 10.1186/s40658-022-00487-7.

Abstract

BACKGROUND

PET nuclides can have a considerable influence on the spatial resolution and image quality of PET/CT scans, which can influence diagnostics in oncology, for example. The individual impact of the positron energy of F, Ga, and Cu on spatial resolution and image quality was compared for PET/CT scans acquired using a clinical, digital scanner.

METHODS

A Jaszczak phantom and a NEMA PET body phantom were filled with F-FDG, Ga-HCl, or Cu-HCl, and PET/CT scans were performed on a Siemens Biograph Vision. Acquired images were analyzed regarding spatial resolution and image quality (recovery coefficients (RC), coefficient of variation within the background, contrast recovery coefficient (CRC), contrast-noise ratio (CNR), and relative count error in the lung insert). Data were compared between scans with different nuclides.

RESULTS

We found that image quality was comparable between F-FDG and Cu-HCl PET/CT measurements featuring similar maximal endpoint energies of the positrons. In comparison, RC, CRC, and CNR were degraded in Ga-HCl data despite similar count rates. In particular, the two smallest spheres of 10 mm and 13 mm diameter revealed lower RC, CRC, and CNR values. The spatial resolution was similar between F-FDG and Cu-HCl but up to 18% and 23% worse compared with PET/CT images of the NEMA PET body phantom filled with Ga-HCl.

CONCLUSIONS

The positron energy of the PET nuclide influences the spatial resolution and image quality of a digital PET/CT scan. The image quality and spatial resolution of Ga-HCl PET/CT images were worse than those of F-FDG or Cu-HCl despite similar count rates.

摘要

背景

正电子发射断层显像(PET)核素可对PET/CT扫描的空间分辨率和图像质量产生相当大的影响,例如这可能会影响肿瘤学诊断。使用临床数字扫描仪获取PET/CT扫描图像,比较了氟(F)、镓(Ga)和铜(Cu)的正电子能量对空间分辨率和图像质量的个体影响。

方法

将F-FDG、Ga-HCl或Cu-HCl注入Jaszczak体模和NEMA PET体模中,并在西门子Biograph Vision上进行PET/CT扫描。对获取的图像进行空间分辨率和图像质量分析(恢复系数(RC)、本底内变异系数、对比恢复系数(CRC)、对比噪声比(CNR)以及肺插入物中的相对计数误差)。比较不同核素扫描之间的数据。

结果

我们发现,在正电子最大终点能量相似的F-FDG和Cu-HCl PET/CT测量中,图像质量相当。相比之下,尽管计数率相似,但Ga-HCl数据中的RC、CRC和CNR有所下降。特别是,直径为10毫米和13毫米的两个最小球体显示出较低的RC、CRC和CNR值。F-FDG和Cu-HCl之间的空间分辨率相似,但与填充Ga-HCl的NEMA PET体模的PET/CT图像相比,分别差高达18%和23%。

结论

PET核素的正电子能量会影响数字PET/CT扫描的空间分辨率和图像质量。尽管计数率相似,但Ga-HCl PET/CT图像的图像质量和空间分辨率比F-FDG或Cu-HCl的要差。