Wolfe Eye Clinic, Marshalltown, IA, USA.
J Refract Surg. 2010 Nov;26(11):851-7. doi: 10.3928/1081597X-20100114-07. Epub 2010 Jan 15.
To report the incidence and management of intraoperative complications while using the IntraLase FS femtosecond laser (Abbott Medical Optics) for flap creation during LASIK.
A prospective log of intraoperative flap creation complications using the femtosecond laser was maintained over 3009 consecutive LASIK surgeries from August 2002 through July 2009.
Eleven (0.37%) intraoperative complications occurred. Eight suction breaks occurred, seven of which were treated with reapplication of the laser, the other by completing the side-cut with surgical dissection. One case of incomplete flap creation was treated with surgical dissection. Two cases of adherent flap were treated with reapplication of laser energy and surgical dissection. All surgeries were completed successfully during the same operative session.
Intraoperative complications during flap creation are minor, infrequent, and can be managed effectively within the same surgical session.
报告使用 IntraLase FS 飞秒激光( Abbott Medical Optics)在 LASIK 中制作角膜瓣时发生的术中并发症及处理方法。
对 2002 年 8 月至 2009 年 7 月 3009 例连续 LASIK 手术中使用飞秒激光制作角膜瓣的术中并发症进行前瞻性记录。
术中发生 11 例(0.37%)并发症。8 例发生负压环中断,其中 7 例经再次应用激光治疗,另 1 例通过手术切开完成侧切口。1 例角膜瓣制作不完全,经手术切开处理。2 例角膜瓣粘连,经再次应用激光和手术切开处理。所有手术均在同一手术期成功完成。
角膜瓣制作过程中的术中并发症轻微、罕见,可在同一手术期内有效处理。