Performance analysis of the OneTouch UltraVue blood glucose monitoring system.

作者信息

Chang Anna, Orth Alice, Le Bryan, Menchavez Perla, Miller Lupe

机构信息

John Muir Physician Network Clinical Research Center, Concord, California 94520, USA.

出版信息

J Diabetes Sci Technol. 2009 Sep 1;3(5):1158-65. doi: 10.1177/193229680900300521.

Abstract

BACKGROUND

OneTouch UltraVue is a new meter for self-monitoring of blood glucose that includes a color display, used-strip ejector, and no-button interface. The system uses an electrochemical biosensor technology based on glucose oxidase chemistry to detect glucose concentrations from 20 to 600 mg/dl (1.1 to 33.3 mmol/liter).

METHODS

Accuracy and reproducibility were evaluated over a wide range of glucose concentrations according to standard criteria. Clinical accuracy was assessed by health care providers (HCPs) in two studies and by diabetes patients in the second study. Reference glucose levels were determined by a YSI 2300 analyzer. Same-day reproducibility and day-to-day reproducibility were also evaluated.

RESULTS

In the accuracy studies, 99.7% and 98.7% of tests by HCPs and 97.0% of tests by patients were within +/-15 mg/dl (+/-0.8 mmol/liter) of the YSI reference for blood glucose <75 mg/dl (<4.2 mmol/liter), and within +/-20% for blood glucose > or =75 mg/dl (> or =4.2 mmol/liter), respectively. Consensus error grid analysis showed that 99.7% and 95.3% of tests by HCPs and 97.0% of tests by patients fell within zone A (i.e., has no effect on clinical action); all other results were in zone B (i.e., altered clinical action, little or no effect on clinical outcome). In the reproducibility studies, the standard deviation was <1.5 mg/dl (<0.1 mmol/liter) for glucose concentrations <100 mg/dl (<5.6 mmol/liter), and the coefficient of variation was <2% for concentrations > or = 100 mg/dl (> or =5.6 mmol/liter).

CONCLUSIONS

OneTouch UltraVue meets standard acceptability criteria for accuracy and reproducibility across a wide range of glucose concentrations. Its simple interface and lack of contact with used strips make it a viable option for older patients and their caregivers.

摘要

背景

OneTouch UltraVue是一款用于自我血糖监测的新型血糖仪,具有彩色显示屏、用过试纸弹出装置和无按钮界面。该系统采用基于葡萄糖氧化酶化学的电化学生物传感器技术,可检测20至600毫克/分升(1.1至33.3毫摩尔/升)的葡萄糖浓度。

方法

根据标准标准,在广泛的葡萄糖浓度范围内评估准确性和可重复性。在两项研究中,由医疗保健提供者(HCP)评估临床准确性,在第二项研究中由糖尿病患者评估。通过YSI 2300分析仪测定参考葡萄糖水平。还评估了同日可重复性和日间可重复性。

结果

在准确性研究中,对于血糖<75毫克/分升(<4.2毫摩尔/升),HCP进行的测试中有99.7%和98.7%以及患者进行的测试中有97.0%在YSI血糖参考值的±15毫克/分升(±0.8毫摩尔/升)范围内;对于血糖≥75毫克/分升(≥4.2毫摩尔/升),分别在±20%范围内。一致性误差网格分析显示,HCP进行的测试中有99.7%和95.3%以及患者进行的测试中有97.0%落在A区(即对临床行动无影响);所有其他结果在B区(即改变临床行动,对临床结果影响很小或无影响)。在可重复性研究中,对于葡萄糖浓度<100毫克/分升(<5.6毫摩尔/升),标准差<1.5毫克/分升(<0.1毫摩尔/升),对于浓度≥100毫克/分升(≥5.6毫摩尔/升),变异系数<2%。

结论

OneTouch UltraVue在广泛的葡萄糖浓度范围内符合准确性和可重复性的标准可接受标准。其简单的界面以及与用过试纸无接触的特点使其成为老年患者及其护理人员的可行选择。

相似文献

1
Performance analysis of the OneTouch UltraVue blood glucose monitoring system.
J Diabetes Sci Technol. 2009 Sep 1;3(5):1158-65. doi: 10.1177/193229680900300521.
2
Analysis article on the performance analysis of the OneTouch UltraVue blood glucose monitoring system.
J Diabetes Sci Technol. 2009 Sep 1;3(5):1166-7. doi: 10.1177/193229680900300522.
3
Precision, accuracy, and user acceptance of the OneTouch SelectSimple blood glucose monitoring system.
J Diabetes Sci Technol. 2011 Nov 1;5(6):1602-9. doi: 10.1177/193229681100500638.
4
Clinical evaluation of a novel test strip technology for blood glucose monitoring: accuracy at hypoglycaemic glucose levels.
Diabetes Res Clin Pract. 2012 Dec;98(3):430-5. doi: 10.1016/j.diabres.2012.09.027. Epub 2012 Oct 9.
6
7
Performance of the CONTOUR® TS Blood Glucose Monitoring System.
J Diabetes Sci Technol. 2011 Jan 1;5(1):198-205. doi: 10.1177/193229681100500128.
8
A human pilot study of the fluorescence affinity sensor for continuous glucose monitoring in diabetes.
J Diabetes Sci Technol. 2012 Mar 1;6(2):362-70. doi: 10.1177/193229681200600222.

引用本文的文献

2
Fabrication of functionalized carbon nanotube buckypaper electrodes for application in glucose biosensors.
Biosensors (Basel). 2014 Nov 11;4(4):449-60. doi: 10.3390/bios4040449. eCollection 2014 Dec.
3
Analysis article on the performance analysis of the OneTouch UltraVue blood glucose monitoring system.
J Diabetes Sci Technol. 2009 Sep 1;3(5):1166-7. doi: 10.1177/193229680900300522.

本文引用的文献

1
Performance evaluation of a new blood glucose monitor that requires no coding: the OneTouch Vita System.
J Diabetes Sci Technol. 2008 Sep;2(5):814-8. doi: 10.1177/193229680800200510.
2
Self-monitoring of blood glucose in type 2 diabetes: a new look at published trials.
Diabetologia. 2008 Apr;51(4):686-8. doi: 10.1007/s00125-008-0946-7. Epub 2008 Feb 20.
3
Standards of medical care in diabetes--2008.
Diabetes Care. 2008 Jan;31 Suppl 1:S12-54. doi: 10.2337/dc08-S012.
5
Assessing glucose meter accuracy.
Curr Med Res Opin. 2006 Nov;22(11):2167-74. doi: 10.1185/030079906X148274.
7
Global prevalence of diabetes: estimates for the year 2000 and projections for 2030.
Diabetes Care. 2004 May;27(5):1047-53. doi: 10.2337/diacare.27.5.1047.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验