Stiffness and endurance of a locking compression plate fixed on fractured femur.

作者信息

Kanchanomai Chaosuan, Muanjan Panurungsit, Phiphobmongkol Vajara

机构信息

Department of Mechanical Engineering, Faculty of Engineering, Thammasat University, Klong-Luang, Pathumthani, Thailand.

出版信息

J Appl Biomech. 2010 Feb;26(1):10-6. doi: 10.1123/jab.26.1.10.

Abstract

The effects of locking screw position (long column fixation--long distance between the nearest screws to the fracture--and short column fixation--short distance between the nearest screws to the fracture) and fracture gap size (1-mm and 8-mm transverse fracture gap) on stiffness and fatigue of fractured femur fixed with a locking compression plate (LCP) were biomechanically evaluated. The stiffness of 1-mm fracture gap models and that of intact femoral model were in the range of 270-284 N/mm, while those of 8-mm fracture gap models were significantly lower (155-170 N/mm). After 1,000,000 cycles of loading, no fracture of LCP of 1-mm fracture gap models fixed in either long column or short column fashions occurred. On the other hand, the complete fractures of LCPs of 8-mm fracture gap models fixed in long column and short column fashions occurred after 51,500 and 42,000 cycles of loading, respectively. These results suggest that the full weight loading may be allowed for the patient with 1-mm transverse femoral fracture fixed with an LCP. On the other hand, the full load of walking should be avoided for the patient with 8-mm transverse femoral fracture fixed with an LCP before adequate healing.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验