Heath P J, Marks L F
Shackleton Department of Anaesthetics, Southampton General Hospital.
Anaesthesia. 1991 Mar;46(3):213-6. doi: 10.1111/j.1365-2044.1991.tb09412.x.
Undetected defects in the inner tube of the Bain coaxial anaesthetic breathing system may result in a greatly increased apparatus deadspace. Several authorities have advocated tests intended to detect inner tube problems; however, the efficacy of these tests has never been validated. In this study none of the tests were able to detect all the induced defects. A modification of an existing test using the backbar pressure-relief valve and a new double occlusion test were sufficiently sensitive to detect all defects.