Evaluation of a rebound tonometer (Tonovet) in clinically normal cat eyes.

作者信息

Rusanen Elina, Florin Marion, Hässig Michael, Spiess Bernhard M

机构信息

Department of Equine and Small Animal Medicine, Section of Surgery, University of Helsinki, Helsinki, Finland.

出版信息

Vet Ophthalmol. 2010 Jan;13(1):31-6. doi: 10.1111/j.1463-5224.2009.00752.x.

Abstract

OBJECTIVE

To determine the accuracy of and to establish reference values for a rebound tonometer (Tonovet) in normal feline eyes, to compare it with an applanation tonometer (Tonopen Vet) and to evaluate the effect of topical anesthesia on rebound tonometry.

PROCEDURES

Six enucleated eyes were used to compare both tonometers with direct manometry. Intraocular pressure (IOP) was measured in 100 cats to establish reference values for rebound tonometry. Of these, 22 cats were used to compare rebound tonometry with and without topical anesthesia and 33 cats to compare the rebound and applanation tonometers. All evaluated eyes were free of ocular disease.

RESULTS

Both tonometers correlated well with direct manometry. The best agreement with the rebound tonometer was achieved between 25-50 mmHg. The applanation tonometer was accurate at pressures between 0 and 30 mmHg. The mean IOP in clinically normal cats was 20.74 mmHg with the rebound tonometer and 18.4 mmHg with the applanation tonometer. Topical anesthesia did not significantly affect rebound tonometry.

CONCLUSIONS

As the rebound tonometer correlated well with direct manometry in the clinically important pressure range and was well tolerated by cats, it appears suitable for glaucoma diagnosis. The mean IOP obtained with the rebound tonometer was 2-3 mmHg higher than that measured with the applanation tonometer. This difference is within clinically acceptable limits, but indicates that the same type of tonometer should be used in follow-up examinations in a given cat.

摘要

目的

确定回弹式眼压计(Tonovet)在正常猫眼中的准确性并建立参考值,将其与压平式眼压计(Tonopen Vet)进行比较,并评估表面麻醉对回弹式眼压测量的影响。

方法

使用6只摘除的眼球将两种眼压计与直接测压法进行比较。对100只猫测量眼内压(IOP)以建立回弹式眼压测量的参考值。其中,22只猫用于比较表面麻醉和未使用表面麻醉时的回弹式眼压测量,33只猫用于比较回弹式眼压计和压平式眼压计。所有评估的眼睛均无眼部疾病。

结果

两种眼压计与直接测压法的相关性都很好。回弹式眼压计在25 - 50 mmHg之间与直接测压法的一致性最佳。压平式眼压计在0至30 mmHg的压力下是准确的。使用回弹式眼压计测量,临床正常猫的平均IOP为20.74 mmHg,使用压平式眼压计测量为18.4 mmHg。表面麻醉对回弹式眼压测量没有显著影响。

结论

由于回弹式眼压计在临床重要的压力范围内与直接测压法相关性良好,且猫对其耐受性良好,因此它似乎适用于青光眼诊断。使用回弹式眼压计获得的平均IOP比使用压平式眼压计测量的高2 - 3 mmHg。这种差异在临床可接受范围内,但表明在对同一只猫进行后续检查时应使用同一类型的眼压计。