To compare and evaluate the accuracy of intraocular pressure (IOP) measured through a therapeutic contact lens, using applanation (TonoPen XL(®)) and rebound (TonoVet(®)) tonometers in enucleated dog eyes.
A total of 30 enucleated eyes from 15 beagle dogs.
To measure accurate IOP, the anterior chamber of each enucleated eye was cannulated with two 26-gauge needles and two polyethylene tubes were connected vertically to an adjustable reservoir bag of normal saline and a pressure transducer. IOP was measured by the TonoPen XL(®) followed by the TonoVet(®) without a contact lens. After a contact lens was applied to the cornea, IOP was re-measured in the same order. Three consecutive IOP measurements were performed using both tonometers.
Without the contact lens, the IOP values obtained by both tonometers correlated well according to the regression analysis (TonoVet(®): γ(2) = 0.98, TonoPen XL(®): γ(2) = 0.97, P < 0.001). The TonoPen XL(®) consistently underestimated values as transducer IOP increased; however, IOP values measured with the TonoPen XL(®) were in close agreement and were less variable than those determined with the TonoVet(®) when a contact lens was applied to the cornea. Bland-Altman analysis was used to determine the lower and upper limits of agreement (TonoVet(®): -29.7 and +21.1 mmHg, TonoPen XL(®): -3.9 and +3.6 mmHg) between the two devices.
This study suggests that the TonoPen XL(®) is a useful tonometer for dogs wearing therapeutic contact lenses, and importantly, contact lenses would not need to be removed prior to IOP measurement.
使用压平眼压计(TonoPen XL(®))和回弹眼压计(TonoVet(®))对比并评估通过治疗性隐形眼镜在摘除的犬眼上测量眼压(IOP)的准确性。
来自15只比格犬的共30只摘除的眼睛。
为测量准确的眼压,每只摘除的眼睛前房用两根26号针头插管,两根聚乙烯管垂直连接到一个可调节的生理盐水储液袋和一个压力传感器上。先用TonoPen XL(®)测量眼压,然后在不戴隐形眼镜的情况下用TonoVet(®)测量。在角膜上戴上隐形眼镜后,按相同顺序重新测量眼压。使用两种眼压计连续进行三次眼压测量。
在不戴隐形眼镜的情况下,根据回归分析,两种眼压计测得的眼压值相关性良好(TonoVet(®):γ(2) = 0.98,TonoPen XL(®):γ(2) = 0.97,P < 0.001)。随着传感器眼压升高,TonoPen XL(®)持续低估眼压值;然而,当在角膜上戴上隐形眼镜时,用TonoPen XL(®)测得的眼压值与用TonoVet(®)测得的值密切一致且变异性更小。采用Bland-Altman分析确定两种设备之间的一致性下限和上限(TonoVet(®):-29.7和+21.1 mmHg,TonoPen XL(®):-3.9和+3.6 mmHg)。
本研究表明,TonoPen XL(®)是一种适用于佩戴治疗性隐形眼镜犬的眼压计,重要的是,在测量眼压前无需摘除隐形眼镜。