Egan W G, Hilgeman T
Appl Opt. 1975 May 1;14(5):1137-42. doi: 10.1364/AO.14.001137.
A new system for absolute total diffuse transmission and reflection measurements using two integrating spheres is described; it extends the normal spectral range into the uv and ir. Furthermore, as part of the system, a unique self-calibrating technique has been incorporated. One sphere is coated with BaSO(4) for use between wavelengths of 0.185 microm and 2.0 microm, and the other is coated with sulfur flowers for use between 1.5 microm and 12 microm. The BaSO(4) and sulfur fulfill the diffuseness and high reflectance characteristics necessary for an integrating sphere. The feature of the system for calibration of the spheres uses goniophotometric measurements of diffuse transmission and reflection. These reflection intercalibrations are better in the visual range than a fraction of a percent (absolutely); however, transmission intercalibrations on quasi-specular samples may have absolute differences of several percent, thus requiring a goniophotometric transmission measurement as the valid determination.
描述了一种使用两个积分球进行绝对总漫透射和反射测量的新系统;它将正常光谱范围扩展到紫外和红外区域。此外,作为该系统的一部分,还采用了一种独特的自校准技术。一个积分球涂有硫酸钡,用于0.185微米至2.0微米波长范围,另一个涂有硫粉,用于1.5微米至12微米波长范围。硫酸钡和硫满足积分球所需的漫射性和高反射率特性。该系统用于积分球校准的特点是采用了漫透射和反射的测角光度测量。在视觉范围内,这些反射相互校准的绝对误差优于百分之几;然而,对准镜面样品的透射相互校准可能存在百分之几的绝对差异,因此需要进行测角光度透射测量作为有效测定。