The investigation of drug penetration into the stratum corneum (SC) by tape-stripping requires an accurate measure of the amount of SC on each tape-strip in order to determine the depth inside the SC. This study applies infrared densitometry (IR-D) to in vitro tape-stripping using the novel Squame Scan(R) 850A. The device had recently been shown to provide accurate measurements of the SC depth for tape-stripping in vivo. Furthermore, the suitability of IR-D for determining the endpoint of tape-stripping, i.e. complete SC removal, was tested. The SC depth was computed from the IR-D data of sequential tape-strips and compared to the results of a protein assay as gold standard. IR-D provided accurate depth results both for freshly excised skin and for skin stored frozen for up to 3 months. In addition, the lower limit of quantification of IR-D indicates the complete removal of the SC (less than 5% of the total SC remaining) and can be used for adjusting the number of tapes applied in situ. Therefore, IR-D is an accurate, fast and non-destructive method for SC depth determination.
胶带撕除法研究药物渗透进入角质层(SC),需要准确测量每条胶带所去除的 SC 量,以确定 SC 的深度。本研究应用近红外密度测定法(IR-D),采用新型 Squame Scan(R) 850A 进行体外胶带撕除法。该设备最近已被证明可用于准确测量体内胶带撕除法的 SC 深度。此外,还测试了 IR-D 确定胶带撕除法终点(即完全去除 SC)的适用性。通过对连续胶带的 IR-D 数据进行计算,得出 SC 深度,并与蛋白质检测法(金标准)的结果进行比较。IR-D 为新鲜切取的皮肤和冷冻保存长达 3 个月的皮肤提供了准确的深度结果。此外,IR-D 的定量下限表明 SC 已完全去除(剩余的 SC 不到总 SC 的 5%),可用于调整原位应用的胶带数量。因此,IR-D 是一种准确、快速且非破坏性的 SC 深度测定方法。