Biopharmaceutics and Pharmaceutical Technology, Saarland University, Saarbrücken, Germany.
Skin Pharmacol Physiol. 2012;25(5):249-56. doi: 10.1159/000339905. Epub 2012 Jul 21.
To perform accurate tape-stripping measurements and to control for site-specific and interindividual differences the amount of stratum corneum (SC) removed by each tape and the total SC thickness must be known. The purpose of this study was to evaluate the use of near-infrared (NIR) densitometry at λ = 850 nm for in situ determination of the total SC thickness. Quantitative tape stripping was performed on pig ear skin. The amount of SC removed by each tape was measured by NIR densitometry and by microprotein assay. Derived from the linear correlation between both measurements, a conversion factor was calculated that relates the individual NIR densitometry readings to the thickness of the SC on the corresponding tape (l(SC-tape) [µm] = (abs.(850) - abs.(850)(blank))/23.9). The total SC thickness was determined based on the accumulated values of all tapes applied in quantitative tape stripping and compared to the values obtained from microscopic cross sections of biopsies. The total SC thickness was correctly determined by infrared densitometry independent of storage time and conditions (4°C up to 24 h; -21°C up to 3 months) in comparison with the standard histological evaluation.
为了进行准确的胶带剥离测量,并控制特定部位和个体间的差异,必须知道每段胶带去除的角质层(SC)的量和总 SC 厚度。本研究旨在评估使用近红外(NIR)分光光度法在λ=850nm 处原位测定总 SC 厚度。在猪耳皮肤上进行定量胶带剥离。用 NIR 分光光度法和微量蛋白测定法测量每段胶带去除的 SC 量。从这两种测量的线性相关中,计算出一个转换因子,将个体 NIR 密度读数与相应胶带上 SC 的厚度联系起来(l(SC-tape)[µm]=(abs.(850)-abs.(850)(空白))/23.9)。基于定量胶带剥离中应用的所有胶带的累积值来确定总 SC 厚度,并与从活检的显微镜横截面获得的值进行比较。与标准组织学评估相比,红外分光光度法可以正确地确定总 SC 厚度,不受储存时间和条件(4°C 至 24 小时;-21°C 至 3 个月)的影响。