应用频域光学相干断层扫描仪/视野计的重复测量功能对正常眼和青光眼患者的视网膜神经纤维层厚度进行重复性测量。

Reproducibility of retinal nerve fiber thickness measurements using the test-retest function of spectral OCT/SLO in normal and glaucomatous eyes.

机构信息

Department of Ophthalmology, Seoul National University College of Medicine, Korea.

出版信息

J Glaucoma. 2010 Dec;19(9):637-42. doi: 10.1097/IJG.0b013e3181ca7cbe.

Abstract

PURPOSE

To evaluate the reproducibility of retinal nerve fiber layer (RNFL) thickness measurements in normal and glaucoma patients using the spectral optical coherence tomography (OCT)/scanning laser ophthalmosope (SLO).

METHODS

One experienced operator measured peripapillary RNFL thickness 3 times in succession in 98 normal participants and 79 glaucoma patients by using the test-retest function of the spectral OCT/SLO. The intraclass correlation coefficient (ICC), coefficient of variation, and test-retest variability (TRV) were analyzed for the mean RNFL thickness, and the RNFL thickness in each quadrant and at each clock-hour, in randomly selected eyes. The correlation coefficient was calculated between the TRV and the mean RNFL thickness, and between the TRV and the mean deviation of the Humphrey visual field.

RESULTS

The ICC for mean RNFL thickness was 0.988 in the normal group and 0.993 in the glaucoma group. The ICC for RNFL thickness in all quadrants and clock-hours in both normal and glaucoma patients ranged from 0.951 to 0.992. The coefficient of variation was between 1.9% and 5.0% in the normal group and between 2.0% and 5.3% in the glaucoma group. The TRV was between 4.3 and 11.5 μm in the normal group and between 3.8 and 10.0 μm in the glaucoma group. The TRV showed no correlation with the mean RNFL thickness or the mean deviation (Pearson r>0.15, P>0.2).

CONCLUSIONS

Spectral OCT/SLO equipped with a test-retest function showed excellent reproducibility with regard to RNFL thickness measurements. The severity of glaucoma did not affect the reproducibility. These findings suggest that spectral OCT/SLO fulfills one of the basic premises for the diagnosis and progression analysis of glaucoma.

摘要

目的

利用频域光学相干断层扫描(OCT)/扫描激光检眼镜(SLO)评估正常人和青光眼患者视网膜神经纤维层(RNFL)厚度测量的可重复性。

方法

一位经验丰富的操作者使用频域 OCT/SLO 的测试-复测功能,对 98 名正常参与者和 79 名青光眼患者的视盘周围 RNFL 厚度进行了 3 次连续测量。分析了随机选择的眼的平均 RNFL 厚度以及每个象限和每个时钟小时的 RNFL 厚度的组内相关系数(ICC)、变异系数和测试-复测变异性(TRV)。计算了 TRV 与平均 RNFL 厚度之间以及与 Humphrey 视野平均偏差之间的相关系数。

结果

正常组的平均 RNFL 厚度的 ICC 为 0.988,青光眼组为 0.993。正常人和青光眼患者所有象限和时钟小时的 RNFL 厚度的 ICC 范围为 0.951 至 0.992。正常组的变异系数为 1.9%至 5.0%,青光眼组为 2.0%至 5.3%。正常组的 TRV 为 4.3 至 11.5μm,青光眼组为 3.8 至 10.0μm。TRV 与平均 RNFL 厚度或平均偏差之间无相关性(Pearson r>0.15,P>0.2)。

结论

配备测试-复测功能的频域 OCT/SLO 对 RNFL 厚度测量具有极好的可重复性。青光眼的严重程度并不影响其可重复性。这些发现表明,频域 OCT/SLO 满足了青光眼诊断和进展分析的基本前提之一。