计算机辅助测听与手动测听。

Computer-assisted audiometry versus manual audiometry.

机构信息

Department of Otolaryngology, Sunderland Royal Hospital, Sunderland, UK.

出版信息

Otol Neurotol. 2009 Oct;30(7):876-83. doi: 10.1097/MAO.0b013e3181b120d0.

Abstract

OBJECTIVE

The Otogram is an automated computer-assisted audiometer that allows patients to self-administer audiometry for their pure-tone audiogram. There has been no research published in a peer-reviewed journal, validating its use in an otology outpatient clinic. We therefore proposed to investigate and compare the inter-rater and intrarater accuracy and reliability of audiologists and of the Otogram in an English-speaking British population.

DESIGN

Prospective nonrandomized validation study.

SETTING

Secondary otolaryngology center and otology outpatient clinic.

PARTICIPANTS

Forty-eight NHS patients referred to an otology outpatient clinic.

MAIN OUTCOME MEASURES

Each patient had 2 pure-tone audiograms. Hearing thresholds in decibels hearing level were ascertained by fully trained British audiologists and by the Otogram.

RESULTS

Using the weighted kappa statistic, the level of agreement in air-conduction (kappa = 0.965) and bone-conduction (kappa = 0.927) thresholds between the audiologist and the Otogram on the same patient was equivalent to the inter-rater level of agreement between pairs of audiologists. Approximately 94% of air-conduction thresholds and 91% of bone-conduction thresholds measured by the Otogram fell within 10 dB of thresholds measured by an audiologist. Intrarater comparisons between audiologists were very good for air-conduction (kappa = 0.978) and bone-conduction (kappa = 0.964). The intrarater level of agreement between repeated Otogram thresholds was just as good for air-conduction (kappa = 0.974) and bone-conduction (kappa = 0.945) thresholds.

CONCLUSION

The Otogram is just as reliable as audiologists at determining hearing thresholds. We recommend that the Otogram can be safely used in a controlled clinical setting supervised by audiologists.

摘要

目的

Otogram 是一种自动化计算机辅助听力计,允许患者自行进行纯音听力测试以获得其纯音听力图。目前,还没有在同行评议的期刊上发表过验证其在耳科学门诊中使用的研究。因此,我们提出在讲英语的英国人群中调查和比较听力学家和 Otogram 的组内和组间准确性和可靠性。

设计

前瞻性非随机验证研究。

设置

二级耳鼻喉中心和耳科门诊。

参与者

48 名 NHS 患者被转诊到耳科门诊。

主要观察指标

每位患者均进行 2 次纯音听力测试。由经过充分培训的英国听力学家和 Otogram 确定听力阈值(以分贝听力级表示)。

结果

使用加权 Kappa 统计量,听力学家和 Otogram 在同一患者的气导(kappa=0.965)和骨导(kappa=0.927)阈值之间的一致性水平与听力学家之间的组内一致性水平相当。Otogram 测量的约 94%的气导阈值和 91%的骨导阈值与听力学家测量的阈值相差 10dB 以内。听力学家之间的组内比较在气导(kappa=0.978)和骨导(kappa=0.964)方面非常好。听力学家之间重复 Otogram 阈值的组内一致性在气导(kappa=0.974)和骨导(kappa=0.945)方面也同样良好。

结论

Otogram 在确定听力阈值方面与听力学家一样可靠。我们建议在由听力学家监督的受控临床环境中可以安全使用 Otogram。