Computer-based audiometry allows for novel applications, including remote testing and automation, that may improve the accessibility and efficiency of hearing assessment in various clinical and occupational health settings. This study describes the validity of computer-based, diagnostic air and forehead bone conduction audiometry when compared wtih conventional industry standard audiometry in a sound booth environment. A sample of 30 subjects (19 to 77 years of age) was assessed with computer-based (KUDUwave 5000) and industry standard conventional audiometers (GSI 61) to compare air and bone conduction thresholds and test-retest reliability. Air conduction thresholds for the two audiometers corresponded within 5 dB or less in more than 90% of instances, with an average absolute difference of 3.5 dB (3.8 SD) and a 95% confidence interval of 2.6 to 4.5 dB. Bone conduction thresholds for the two audiometers corresponded within 10 dB or less in 92% of instances, with an average absolute difference of 4.9 dB (4.9 SD) and a 95% confidence interval of 3.6 to 6.1 dB. The average absolute test-retest threshold difference for bone conduction on the industry standard audiometer was 5.1 dB (5.3 SD) and for the computer-based audiometer 7.1 dB (6.4 SD). Computer-based audiometry provided air and bone conduction thresholds within the test-retest reliability limits of industry standard audiometry.

摘要

计算机测听允许新的应用,包括远程测试和自动化,这可能提高听力评估在各种临床和职业健康环境中的可及性和效率。本研究描述了在隔音室内,与传统行业标准测听相比,计算机诊断空气和额骨传导测听的有效性。一个 30 名受试者(19 至 77 岁)的样本,使用计算机(KUDUwave 5000)和行业标准常规测听仪(GSI 61)进行了评估,以比较空气和骨传导阈值和测试-再测试的可靠性。在超过 90%的情况下,两种测听仪的空气传导阈值相差在 5dB 以内,平均绝对差值为 3.5dB(3.8SD),95%置信区间为 2.6 至 4.5dB。在 92%的情况下,两种测听仪的骨传导阈值相差在 10dB 以内,平均绝对差值为 4.9dB(4.9SD),95%置信区间为 3.6 至 6.1dB。行业标准测听仪的骨传导测试-再测试阈值平均绝对差值为 5.1dB(5.3SD),而计算机测听仪的平均绝对差值为 7.1dB(6.4SD)。计算机测听在行业标准测听的测试-再测试可靠性范围内提供了空气和骨传导阈值。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验