光谱域光学相干断层成像中的镜像伪影的临床评估。

Clinical assessment of mirror artifacts in spectral-domain optical coherence tomography.

机构信息

New England Eye Center, Tufts Medical Center, Boston, Massachusetts, USA.

出版信息

Invest Ophthalmol Vis Sci. 2010 Jul;51(7):3714-20. doi: 10.1167/iovs.09-4057. Epub 2010 Feb 24.

Abstract

PURPOSE. To investigate the characteristics of a spectral-domain optical coherence tomography (SD-OCT) image phenomenon known as the mirror artifact, calculate its prevalence, analyze potential risk factors, measure severity, and correlate it to spherical equivalent and central visual acuity (VA). METHODS. OCT macular cube 512 x 128 scans taken between January 2008 and February 2009 at the New England Eye Center were analyzed for the presence of mirror artifacts. Artifact severity was determined by the degree of segmentation breakdown that it caused on the macular map. A retrospective review was conducted of the medical records of patients with artifacts and of a random control group without artifacts. RESULTS. Of 1592 patients, 9.3% (148 patients, 200 eyes) had scans that contained mirror artifacts. A significantly more myopic spherical equivalent (P < 0.001), worse VA (P < 0.001), longer axial lengths (P = 0.004), and higher proportions of moderate to high myopia (P < 0.001) were found in patients with mirror artifacts than in patients without artifacts. Worse VA was associated with increased artifact severity (P = 0.04). CONCLUSIONS. In all scans analyzed, a high prevalence of mirror artifacts was found. This image artifact was often associated with patients with moderate to high myopia. Improvements in instrumentation may be necessary to resolve this problem in moderately and highly myopic eyes. Operators should be advised to properly position the retina when scanning eyes. In cases in which peripheral abnormalities in topographic measurements of retinal thickness are found, corresponding OCT scans should be examined for the presence of mirror artifacts.

摘要

目的。研究一种称为镜像伪影的谱域光相干断层扫描(SD-OCT)图像现象的特征,计算其发生率,分析潜在的危险因素,测量严重程度,并将其与球镜等效和中心视力(VA)相关联。

方法。在 2008 年 1 月至 2009 年 2 月期间,在新英格兰眼科中心进行了 SD-OCT 黄斑立方 512 x 128 扫描,以分析镜像伪影的存在。通过它在黄斑图上引起的分割破裂程度来确定伪影的严重程度。对有伪影的患者的病历和无伪影的随机对照组进行回顾性审查。

结果。在 1592 名患者中,9.3%(148 名患者,200 只眼)的扫描包含镜像伪影。与无伪影的患者相比,有镜像伪影的患者具有更明显的近视球镜等效(P < 0.001)、更差的 VA(P < 0.001)、更长的眼轴长度(P = 0.004)和更高比例的中高度近视(P < 0.001)。VA 越差与伪影严重程度增加相关(P = 0.04)。

结论。在分析的所有扫描中,发现镜像伪影的发生率很高。这种图像伪影通常与中高度近视的患者有关。可能需要改进仪器来解决中度和高度近视眼中的这个问题。操作人员应被告知在扫描眼睛时正确定位视网膜。在发现视网膜厚度的地形图测量存在外周异常的情况下,应检查相应的 OCT 扫描是否存在镜像伪影。

相似文献

引用本文的文献

7
Artifacts in Macular Optical Coherence Tomography.黄斑光学相干断层扫描中的伪像
J Curr Ophthalmol. 2020 Apr 30;32(2):123-131. doi: 10.4103/JOCO.JOCO_83_20. eCollection 2020 Apr-Jun.
9
corneal confocal microscopy aided by optical coherence tomography.光学相干断层扫描辅助角膜共焦显微镜检查
Biomed Opt Express. 2019 Apr 26;10(5):2580-2587. doi: 10.1364/BOE.10.002580. eCollection 2019 May 1.

本文引用的文献

9
State-of-the-art retinal optical coherence tomography.视网膜光学相干断层扫描技术的最新进展。
Prog Retin Eye Res. 2008 Jan;27(1):45-88. doi: 10.1016/j.preteyeres.2007.07.005. Epub 2007 Aug 11.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验