Chen Wei-Rong, Meng Qian-Li, Ye Hai-Yun, Liu Yi-Zhi
State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou 510060, China.
Zhonghua Yan Ke Za Zhi. 2009 Dec;45(12):1084-8.
To evaluate the visual quality after combined implantation of refractive and diffractive multifocal intraocular lenses (IOL) in both eyes.
Prospective, nonrandomized study. Age-related cataract patients were divided into multifocal IOL group (MIOL group) and monofocal IOL group (SIOL group) and received phacoemulsification. In MIOL group, a ReZoom NXG1 IOL and a Tecnis Z M900 IOL were implanted into the dominant eye and the fellow eye, respectively. In SIOL group, Sensar AR40e IOLs were implanted into both eyes. Monocular or binocular test was performed one month after surgery, including wavefront aberration, contrast sensitivity with or without glare and near stereoacuity.
The root mean square of spherical aberration at pupil diameter of 5 mm was 0.058 +/- 0.159 microm in NXG1 eyes and 0.005 +/- 0.169 microm in ZM900 eyes, both of them were significantly lower than 0.408 +/- 0.160 microm achieved in AR40e eyes (F = 11.734; P = 0.001, 0.000). There was no significant difference in the total ocular aberration, contrast sensitivity and high-order aberration among eyes with these different IOLs. The uncorrected near stereoacuity in MIOL group was (45.60 +/- 35.04) seconds of arc, which was significantly better than (110.00 +/- 41.23) seconds of arc achieved in SIOL group (F = 2.923; P = 0.010).
Combined implantation of multifocal ReZoom NXG1 IOL and Tecnis ZM900 IOL provides a good visual quality with reduced spherical aberration and satisfactory near stereoacuity.
评估双眼联合植入折射型和衍射型多焦点人工晶状体(IOL)后的视觉质量。
前瞻性、非随机研究。年龄相关性白内障患者被分为多焦点IOL组(MIOL组)和单焦点IOL组(SIOL组)并接受超声乳化手术。在MIOL组中,分别将一枚ReZoom NXG1 IOL和一枚Tecnis Z M900 IOL植入优势眼和对侧眼。在SIOL组中,双眼均植入Sensar AR40e IOL。术后1个月进行单眼或双眼检查,包括波前像差、有无眩光时的对比敏感度和近立体视锐度。
NXG1眼在瞳孔直径5mm时的球差均方根为0.058±0.159μm,ZM900眼为0.005±0.169μm,两者均显著低于AR40e眼的0.408±0.160μm(F=11.734;P=0.001,0.000)。使用这些不同IOL的眼在总眼像差、对比敏感度和高阶像差方面无显著差异。MIOL组的未矫正近立体视锐度为(45.60±35.04)角秒,显著优于SIOL组的(110.00±41.23)角秒(F=2.923;P=0.010)。
联合植入多焦点ReZoom NXG1 IOL和Tecnis ZM900 IOL可提供良好的视觉质量,球差降低,近立体视锐度满意。