Physiol Meas. 2010 Apr;31(4):N25-33. doi: 10.1088/0967-3334/31/4/N02. Epub 2010 Mar 5.
A test system has been developed that can be used to calibrate and determine the time response, linearity and temperature sensitivity of a fibre optic oxygen sensor. The simple system obviates the need for precision gas standards and the requirement to generate a true square wave step response, which is seldom achievable. The sensor is mounted in a small chamber containing air or a known fraction of oxygen. By means of a computer-controlled switch, the absolute pressure within the chamber can be changed rapidly to a new steady state value. The partial pressure of oxygen changes in direct proportion to the absolute pressure, and so the accuracy and linearity and response time of the PO(2) calibration are limited only by those of the absolute pressure sensor. The temperature sensitivity of a commercial sensor and a means of correction are also described.
已经开发出一种测试系统,可用于校准和确定光纤氧传感器的时间响应、线性度和温度灵敏度。该简单系统无需精密气体标准,也无需产生真正的方波阶跃响应,因为这种响应很少能够实现。传感器安装在一个小室中,小室中充满空气或已知的氧气分数。通过计算机控制的开关,可以迅速将腔室内的绝对压力改变到新的稳态值。氧的分压力与绝对压力成正比变化,因此 PO(2)校准的精度、线性度和响应时间仅受绝对压力传感器的限制。本文还描述了一种商业传感器的温度灵敏度及其校正方法。