Ide Takeshi, Wang Jianhua, Tao Aizhu, Leng Theodore, Kymionis George D, O'Brien Terrence P, Yoo Sonia H
Department of Ophthalmology, Bascom Palmer Eye Institute, University of Miami Miller School of Medicine, Miami, Florida, USA.
Ophthalmic Surg Lasers Imaging. 2010 Mar-Apr;41(2):250-4. doi: 10.3928/15428877-20100303-15.
To develop a prototype three-dimensional anterior segment spectral-domain optical coherence tomography (SD-OCT) device and demonstrate the feasibility of its use in the operating room.
Single-institution interventional case series including six consecutive patients undergoing Descemet's stripping automated endothelial keratoplasty (DSAEK). The prototype anterior segment SD-OCT was used intraoperatively to search for the presence of interface fluid between the host cornea and the DSAEK graft.
Anterior segment SD-OCT was successfully used intraoperatively during DSAEK. After the initial placement of the graft, no fluid was clinically apparent; however, interface fluid was identified by anterior segment SD-OCT in two of the six cases. After additional aspiration, all patients were fluid-free on follow-up anterior segment SD-OCT scanning.
Intraoperative anterior segment SD-OCT was used successfully to find interface fluid that was clinically undetectable under the microscope. As such, all patients were able to leave the operating room with a fully attached graft.
研发一款原型三维眼前节谱域光学相干断层扫描(SD-OCT)设备,并证明其在手术室使用的可行性。
单机构介入病例系列研究,纳入连续6例接受Descemet膜剥离自动内皮角膜移植术(DSAEK)的患者。术中使用原型眼前节SD-OCT来寻找宿主角膜与DSAEK移植物之间界面液的存在情况。
在DSAEK手术过程中成功术中使用眼前节SD-OCT。最初植入移植物后,临床上未发现明显液体;然而,6例中有2例通过眼前节SD-OCT发现了界面液。经过额外抽吸后,所有患者在随访的眼前节SD-OCT扫描中均无液体。
术中成功使用眼前节SD-OCT发现了显微镜下临床无法检测到的界面液。因此,所有患者均能带着完全贴合的移植物离开手术室。