Abstract

PURPOSE

To evaluate the factors affecting the sensitivity and specificity of the stereometric optic nerve head (ONH) parameters of the Heidelberg Retina Tomograph (HRT) to glaucomatous progression in stereoscopic ONH photographs.

METHODS

The factors affecting the sensitivity and specificity of the vertical cup : disc ratio, the cup : disc area ratio, the cup volume, the rim area and a linear discriminant function to progression were analysed. These parameters were the best indicators of progression in a retrospective study of 476 eyes. The reference standard for progression was the masked evaluation of stereoscopic ONH photographs.

RESULTS

The factors having the most significant effect on the sensitivity and specificity of the stereometric ONH parameters were the reference height difference and the mean topography standard deviation (TSD), indicating image quality. Also, the change in the TSD and age showed consistent, but variably significant, influence on all parameters tested. The sensitivity and specificity improved when there was little change in the reference height, the image quality was good and stable, and the patients were younger. The sensitivity and specificity of the vertical cup : disc ratio was improved by a large disc area and high baseline cup : disc area ratio. The rim area showed a better sensitivity and specificity for progression with a small disc area and low baseline cup : disc area ratio.

CONCLUSION

The factors affecting the sensitivity and specificity of the stereometric ONH parameters to glaucomatous progression in disc photographs are essentially the same as those affecting the measurement variability of the HRT.

摘要

目的

评估海德堡视网膜断层扫描仪(HRT)视神经头(ONH)立体视标参数的立体视标 ONH 照片中青光眼进展的敏感性和特异性的影响因素。

方法

分析影响垂直杯盘比、杯盘面积比、杯容积、边缘面积和线性判别函数对进展的敏感性和特异性的因素。这些参数是对 476 只眼进行回顾性研究的最佳进展指标。进展的参考标准是立体视标 ONH 照片的盲法评估。

结果

对立体视标 ONH 参数的敏感性和特异性影响最大的因素是参考高度差和平均地形图标准差(TSD),这表明了图像质量。此外,TSD 的变化和年龄对所有测试参数均有一致但可变的显著影响。当参考高度变化较小、图像质量良好且稳定以及患者较年轻,敏感性和特异性提高。大的盘面积和高的基线杯盘面积比可提高垂直杯盘比的敏感性和特异性。小的盘面积和低的基线杯盘面积比使边缘面积对进展具有更好的敏感性和特异性。

结论

影响立体视标 ONH 参数对盘照片中青光眼进展的敏感性和特异性的因素与影响 HRT 测量变异性的因素基本相同。