To investigate the role of reference height inter-test variability upon the variability of the stereometric parameters.
204 glaucomatous patients underwent a complete ophthalmological exam, including Heidelberg Retina Tomography 3 (HRT-3). The exclusion criteria were optic disc or retinal pathology that might interfere with the detection of glaucoma progression, TSD >30μm. 4 sets of data were taken during the HRT-3 exam for each patient.
RH variability ranged between -198 and 187. Correlation analysis revealed a linear dependence between the inter-test variability of RH and stereometric parameters change. The most powerful correlations were observed for: RNFL Thickness (r=0.756, p<0.001), Rim Area (r=0.662, p<0.001), C/D Area Ratio (r=-0.663, p<0.001). The least correlated were Height Variation Contour (r=0.31) and Cup Shape Measure (r=0.07, p=0.3). When RH variability did not exceed 25μm, the correlations with stereometric parameters change were not statistically significant (for Rim Area, r=0.21, p>0.05, for C/D Area Ratio, r=-0.13, p=0.22, for RNFL Thickness r=0.06, p=0.52).
For values >25μm, the variability of the RH is a major factor determining test/retest variability for RNFL Thickness, Rim Area, C/D Area, Rim Volume and Linear C/D. Inter-test variability of RH <25μm is an important criterion for the clinical relevance of stereometric parameters changes.
研究参考高度测量间变异性对立体测量参数变异性的作用。
204例青光眼患者接受了包括海德堡视网膜断层扫描3(HRT-3)在内的全面眼科检查。排除标准为可能干扰青光眼进展检测的视盘或视网膜病变,TSD>30μm。每位患者在HRT-3检查期间采集4组数据。
参考高度(RH)变异性在-198至187之间。相关性分析显示,RH测量间变异性与立体测量参数变化之间存在线性依赖关系。观察到的最强相关性为:视网膜神经纤维层厚度(RNFL,r=0.756,p<0.001)、视盘边缘面积(r=0.662,p<0.001)、杯/盘面积比(C/D,r=-0.663,p<0.001)。相关性最弱的是高度变化轮廓(r=0.31)和杯状测量(r=0.07,p=0.3)。当RH变异性不超过25μm时,与立体测量参数变化的相关性无统计学意义(视盘边缘面积,r=0.21,p>0.05;杯/盘面积比,r=-0.13,p=0.22;视网膜神经纤维层厚度,r=0.06,p=0.52)。
对于大于25μm的值,RH变异性是决定视网膜神经纤维层厚度、视盘边缘面积、杯/盘面积、视盘边缘体积和线性杯/盘比测量重测变异性的主要因素。RH测量间变异性<25μm是立体测量参数变化临床相关性的重要标准。