使用三种探测器测量 Elekta 放射治疗设备的门控特性。

Gating characteristics of an Elekta radiotherapy treatment unit measured with three types of detector.

机构信息

Joint Physics Department, Institute of Cancer Research and Royal Marsden Hospital, Downs Road, Sutton, Surrey, SM2 5PT, UK.

出版信息

Phys Med Biol. 2010 Apr 21;55(8):N201-10. doi: 10.1088/0031-9155/55/8/N02. Epub 2010 Mar 26.

Abstract

The characteristics of an Elekta Precise treatment machine with a gating interface were investigated. Three detectors were used: a Farmer ionization chamber, a MatriXX ionization chamber array and an in-house, single pulse-measurement ionization chamber (IVC). Measurements were made of dosimetric accuracy, flatness and symmetry characteristics and duty cycle for a range of beam-on times and gating periods. Results were compared with a standard ungated delivery as a reference. For all beam-on times, down to 0.5 s, dosimetric differences were below +/-1% and flatness and symmetry parameter variations were below +/-1.5%. For the shorter beam-on times the in-house detector deviated from the other two detectors, suggesting that this device should be used in conjunction with other detectors for absolute dosimetry purposes. However, it was found to be useful for studying gated beam characteristics pulse by pulse.

摘要

研究了具有门控接口的 Elekta Precise 治疗机的特性。使用了三种探测器:一个 Farmer 电离室、一个 MatriXX 电离室阵列和一个内部的单脉冲测量电离室(IVC)。测量了剂量准确性、平坦度和对称性特征以及一系列波束开启时间和门控周期的占空比。结果与标准无门控输送作为参考进行了比较。对于所有的波束开启时间,低至 0.5 秒,剂量差异低于 +/-1%,平坦度和对称性参数变化低于 +/-1.5%。对于较短的波束开启时间,内部探测器与其他两个探测器存在偏差,这表明该设备应与其他探测器结合使用,用于绝对剂量测量目的。然而,它被发现对于逐脉冲研究门控波束特性非常有用。