Elsie Widdowson Laboratory, MRC Human Nutrition Research, Cambridge, UK.
Obesity (Silver Spring). 2010 Dec;18(12):2385-91. doi: 10.1038/oby.2010.71. Epub 2010 Apr 1.
Abdominal fat, and in particular, visceral adipose tissue (VAT), is the critical fat depot associated with metabolic aberrations. At present, VAT can only be accurately measured by computed tomography or magnetic resonance imaging (MRI). This study was designed to compare a new abdominal bioelectrical impedance (BIA) device against total abdominal adipose tissue (TAAT) and VAT area measurements made from an abdominal MRI scan, and to assess its reliability and accuracy. One-hundred twenty participants were recruited, stratified by gender and BMI. Participants had triplicate measures of abdominal fat and waist circumference (WC) with the AB-140 (Tanita, Tokyo, Japan) and WC measurements using a manual tape measure. A single abdominal MRI scan was performed as the reference method. Triplicate measures with the AB-140 showed excellent precision for "visceral fat level," trunk fat %, and WC. AB-140 "visceral fat level" showed significantly stronger correlations with MRI TAAT area than with MRI VAT area (r = 0.94 vs. 0.65 in men and 0.92 vs. 0.64 in women). AB-140 WC showed good correlation with manual WC measurements (r = 0.95 in men and 0.90 in women). AB-140 and manual WCs showed comparable correlations with MRI TAAT area (r = 0.92 and 0.96 in men and 0.88 and 0.88 in women). AB-140 is a simple, quick, and precise technique to measure abdominal fat and WC in healthy adults. It provides a useful proxy for TAAT measured by MRI, comparable to the correlation obtained with manual WC measurements. Neither the AB-140 abdominal fat measures nor WC measurement appear to provide a useful proxy measure of VAT.
腹部脂肪,尤其是内脏脂肪组织 (VAT),是与代谢异常相关的关键脂肪库。目前,只有通过计算机断层扫描或磁共振成像 (MRI) 才能准确测量 VAT。本研究旨在比较一种新的腹部生物电阻抗 (BIA) 设备与来自腹部 MRI 扫描的总腹部脂肪 (TAAT) 和 VAT 面积测量值,并评估其可靠性和准确性。120 名参与者按性别和 BMI 分层招募。参与者使用 AB-140(Tanita,东京,日本)进行了三次腹部脂肪和腰围 (WC) 测量,并使用手动卷尺进行了 WC 测量。作为参考方法,对所有参与者进行了一次腹部 MRI 扫描。AB-140 的三次测量对于“内脏脂肪水平”、躯干脂肪 % 和 WC 具有极好的精度。AB-140 的“内脏脂肪水平”与 MRI TAAT 面积的相关性明显强于与 MRI VAT 面积的相关性(男性为 0.94 与 0.65,女性为 0.92 与 0.64)。AB-140 WC 与手动 WC 测量具有良好的相关性(男性为 0.95,女性为 0.90)。AB-140 和手动 WC 与 MRI TAAT 面积的相关性相当(男性为 0.92 和 0.96,女性为 0.88 和 0.88)。AB-140 是一种简单、快速、精确的技术,可用于测量健康成年人的腹部脂肪和 WC。它提供了一个有用的 MRI 测量 TAAT 的替代指标,与手动 WC 测量获得的相关性相当。AB-140 的腹部脂肪测量值和 WC 测量值似乎都不能作为 VAT 的有用替代指标。