Med Phys. 2010 Mar;37(3):1273-84. doi: 10.1118/1.3301596.
Accurate measurement of the three-dimensional (3D) rigid body and surface kinematics of the natural human knee is essential for many clinical applications. Existing techniques are limited either in their accuracy or lack more realistic experimental evaluation of the measurement errors. The purposes of the study were to develop a volumetric model-based 2D to 3D registration method, called the weighted edge-matching score (WEMS) method, for measuring natural knee kinematics with single-plane fluoroscopy to determine experimentally the measurement errors and to compare its performance with that of pattern intensity (PI) and gradient difference (GD) methods.
The WEMS method gives higher priority to matching of longer edges of the digitally reconstructed radiograph and fluoroscopic images. The measurement errors of the methods were evaluated based on a human cadaveric knee at 11 flexion positions.
The accuracy of the WEMS method was determined experimentally to be less than 0.77 mm for the in-plane translations, 3.06 mm for out-of-plane translation, and 1.13 degrees for all rotations, which is better than that of the PI and GD methods.
A new volumetric model-based 2D to 3D registration method has been developed for measuring 3D in vivo kinematics of natural knee joints with single-plane fluoroscopy. With the equipment used in the current study, the accuracy of the WEMS method is considered acceptable for the measurement of the 3D kinematics of the natural knee in clinical applications.
准确测量自然膝关节的三维(3D)刚体和表面运动学对于许多临床应用至关重要。现有的技术要么在准确性上受到限制,要么缺乏对测量误差的更现实的实验评估。本研究的目的是开发一种基于体积模型的 2D 到 3D 配准方法,称为加权边缘匹配得分(WEMS)方法,用于通过单平面荧光透视术测量自然膝关节运动学,以确定测量误差,并将其性能与模式强度(PI)和梯度差(GD)方法进行比较。
WEMS 方法优先匹配数字重建射线照片和荧光透视图像的较长边缘。基于 11 个屈曲位置的人体尸体膝关节评估了这些方法的测量误差。
实验确定 WEMS 方法的精度对于平面内平移小于 0.77mm,对于平面外平移为 3.06mm,对于所有旋转为 1.13 度,优于 PI 和 GD 方法。
已经开发了一种新的基于体积模型的 2D 到 3D 配准方法,用于通过单平面荧光透视术测量自然膝关节的 3D 运动学。在当前研究中使用的设备中,WEMS 方法的精度被认为可接受,可用于临床应用中自然膝关节的 3D 运动学测量。