CEA, DSM, IRAMIS, SIS2M, LSDRM CEA Saclay F-91191 Gif-sur-Yvette, France.
J Magn Reson. 2010 Jul;205(1):75-85. doi: 10.1016/j.jmr.2010.04.003. Epub 2010 Apr 10.
Starting from general results of magnetostatics, we give fundamental considerations on the design and characterization of permanent magnets for NMR based on harmonic analysis and symmetry. We then propose a simple geometry that takes advantage of some of these considerations and discuss the practical aspects of the assembly of a real magnet based on this geometry, involving the characterization of its elements, the optimization of the layout and the correction of residual inhomogeneities due to material and geometry imperfections. We report with this low-cost, light-weight magnet (100 euros and 1.8 kg including the aluminum frame) a field of 120 mT (5.1 MHz proton) with a 10 ppm natural homogeneity over a sphere of 1.5 mm in diameter.
从静磁场的一般结果出发,我们基于谐分析和对称性对基于 NMR 的永磁体的设计和特性进行了基本的考虑。然后,我们提出了一种简单的几何形状,利用了其中的一些考虑因素,并讨论了基于该几何形状组装实际磁铁的实际方面,包括其元件的特性、布局的优化以及由于材料和几何缺陷引起的残余非均匀性的校正。我们报告了这个低成本、轻量级的磁铁(包括铝制框架在内,价格为 100 欧元,重量为 1.8 公斤),其磁场为 120 mT(5.1 MHz 质子),在直径为 1.5 毫米的球体上具有 10 ppm 的自然均匀性。