OBJECTIVES

This in vitro study aimed at evaluating the ability of Laser Fluorescence device (LFpen) in quantifying approximal caries lesions in primary molars.

METHODS

Two examiners assessed 123 approximal surfaces of primary molars using the DIAGNOdent pen (LFpen). Surfaces were determined to be either sound with white-spot lesions or have small cavitations. After sectioning, lesion depth was determined through polarized light microscopy. The intra-/inter-examiner agreement was calculated using intraclass correlation coefficient (ICC) and Bland-Altman analyses. Furthermore, Spearman correlation coefficients (Rs) were calculated between LFpen readings and lesion depth.

RESULTS

Correlation between LFpen values and lesion depth was low for both examiners (Rs=0.36 and 0.51), especially when cavitated lesions were excluded from the analysis (Rs=0.22 and 0.40). For all surfaces, ICC revealed intra- and inter-examiner reproducibility values of 0.75 and 0.63, respectively, but when only non-cavitated surfaces were analyzed, these values decreased (0.41 and 0.33, respectively).

CONCLUSIONS

LFpen readings present low correlation with approximal caries lesion depth and low reproducibility, especially in white-spot lesions. Therefore, the device could not be a suitable method for monitoring non-cavitated approximal caries lesion in primary molars.

摘要

目的

本体外研究旨在评估激光荧光设备(LFpen)定量评估乳磨牙邻面龋损的能力。

方法

两名检查者使用 DIAGNOdent 笔(LFpen)评估了 123 个乳磨牙邻面。根据有无白垩斑或小龋洞,将表面判断为正常或有病变。在切片后,通过偏光显微镜确定病变深度。使用组内相关系数(ICC)和 Bland-Altman 分析评估了内/间检查者一致性。此外,计算了 LFpen 读数与病变深度之间的 Spearman 相关系数(Rs)。

结果

两名检查者的 LFpen 值与病变深度的相关性均较低(Rs=0.36 和 0.51),尤其是排除龋洞病变时(Rs=0.22 和 0.40)。对于所有表面,ICC 显示内/间检查者的重现性值分别为 0.75 和 0.63,但当仅分析无龋洞病变时,这些值降低(分别为 0.41 和 0.33)。

结论

LFpen 读数与邻面龋损深度的相关性较低,重现性也较低,尤其是在白垩斑病变中。因此,该设备可能不是监测乳磨牙非龋性邻面龋损的合适方法。