State Key Laboratory of Transient Optics and Photonics, Xi'an Institute of Optics and Precision Mechanics, Chinese Academy of Sciences, Information road 17, Xi'an, Shaanxi, 710119, China.
Appl Spectrosc. 2010 May;64(5):514-9. doi: 10.1366/000370210791211709.
A highly sensitive microstructured polymer optical fiber (MPOF) probe for hydrogen peroxide was made by forming a rhodamine 6G-doped titanium dioxide film on the side walls of array holes in an MPOF. It was found that hydrogen peroxide only has a response to the MPOF probe in a certain concentration of potassium iodide in sulfuric acid solution. The calibration graph of fluorescence intensity versus hydrogen peroxide concentration is linear in the range of 1.6 x 10(-7) mol/L to 9.6 x 10(-5) mol/L. The method, with high sensitivity and a wide linear range, has been applied to the determination of trace amounts of hydrogen peroxide in a few real samples, such as rain water and contact lens disinfectant, with satisfactory results.
一种高灵敏度的微结构聚合物光纤(MPOF)探针,通过在 MPOF 的阵列孔侧壁上形成罗丹明 6G 掺杂的二氧化钛薄膜来制备。结果发现,过氧化氢仅在硫酸溶液中一定浓度的碘化钾中对 MPOF 探针有响应。荧光强度与过氧化氢浓度的校准曲线在 1.6×10^(-7)mol/L 至 9.6×10^(-5)mol/L 范围内呈线性。该方法具有灵敏度高、线性范围宽的特点,已应用于雨水和隐形眼镜消毒剂等几种实际样品中痕量过氧化氢的测定,结果令人满意。