Suppr超能文献

一种新型高强度聚焦超声睫状体光凝术设备的组织学效应

Histologic effects of a new device for high-intensity focused ultrasound cyclocoagulation.

作者信息

Aptel Florent, Charrel Thomas, Palazzi Xavier, Chapelon Jean-Yves, Denis Philippe, Lafon Cyril

机构信息

INSERM, U556, Lyon, France.

出版信息

Invest Ophthalmol Vis Sci. 2010 Oct;51(10):5092-8. doi: 10.1167/iovs.09-5135. Epub 2010 May 19.

Abstract

PURPOSE

To evaluate the histologic effects and clinical outcomes of high-intensity focused ultrasound (HIFU) delivered by miniaturized annular transducers for ciliary body coagulation in an animal study.

METHODS

Eighteen eyes of 18 rabbits were insonified with a ring comprising a six-sector transducer, to produce a 12.8-mm-diameter circular lesion. Six sectors were activated in six rabbits (group 1), five sectors in six rabbits (group 2), and four sectors in six rabbits (group 3) at 2 W acoustic power. The rabbits were examined before treatment (day 0) and after treatment on days 1, 7, 15, 21, and 28. Detailed qualitative and semiquantitative histopathologic analyses of the enucleated eyes were performed.

RESULTS

In the treated eyes, intraocular pressure changes ranged from -16.6 mm Hg (-55.3%) at day 28 to -8.9 mm Hg (-29.7%) at day 7 in group 1, from -4.7 mm Hg (-25.5%) at day 28 to -1.4 mm Hg (-7.6%) at day 21 in group 2 and from -7.9 mm Hg (-28.1%) at day 28 to +2.0 mm Hg (+7.1%) at day 7 in group 3. No macroscopic abnormalities were observed in the anterior segment or fundus. Histologic examination showed segmental-to-annular lesions in the ciliary processes, caused mainly by coagulation necrosis, whereas the sclera and lens appeared undamaged. Inflammation was very limited.

CONCLUSIONS

Ultrasonic coagulation of the ciliary body with HIFU delivered via a circular miniaturized transducer seemed to be an effective and well-tolerated method of reducing intraocular pressure in an animal study.

摘要

目的

在一项动物研究中,评估通过小型环形换能器进行高强度聚焦超声(HIFU)睫状体凝固术的组织学效应和临床结果。

方法

用一个由六扇区换能器组成的环对18只兔子的18只眼睛进行超声照射,以产生直径为12.8毫米的圆形病变。在6只兔子(第1组)中激活6个扇区,在6只兔子(第2组)中激活5个扇区,在6只兔子(第3组)中激活4个扇区,声功率为2瓦。在治疗前(第0天)以及治疗后第1、7、15、21和28天对兔子进行检查。对摘除的眼睛进行了详细的定性和半定量组织病理学分析。

结果

在治疗的眼睛中,第1组眼压变化范围从第28天的-16.6毫米汞柱(-55.3%)到第7天的-8.9毫米汞柱(-29.7%),第2组从第28天的-4.7毫米汞柱(-25.5%)到第21天的-1.4毫米汞柱(-7.6%),第3组从第28天的-7.9毫米汞柱(-28.1%)到第7天的+2.0毫米汞柱(+7.1%)。在前房或眼底未观察到宏观异常。组织学检查显示睫状突有节段性至环形病变,主要由凝固性坏死引起,而巩膜和晶状体未受损。炎症非常有限。

结论

在动物研究中,通过圆形小型换能器进行HIFU睫状体超声凝固术似乎是一种有效且耐受性良好的降低眼压的方法。