导联完整性警示算法可降低 Sprint Fidelis 起搏感知导线断裂患者中不适当电击的发生率。

Lead integrity alert algorithm decreases inappropriate shocks in patients who have Sprint Fidelis pace-sense conductor fractures.

机构信息

Minneapolis Heart Institute Foundation, Minneapolis, Minnesota 55407, USA.

出版信息

Heart Rhythm. 2010 Aug;7(8):1048-55. doi: 10.1016/j.hrthm.2010.05.015. Epub 2010 May 21.

Abstract

BACKGROUND

The Medtronic Sprint Fidelis high-voltage implantable cardioverter-defibrillator (ICD) lead is prone to fracture. The majority of fractures involve the pace-sense (P/S) conductors and may result in multiple inappropriate shocks. The Medtronic lead integrity alert (LIA) algorithm was designed to improve early detection of transient P/S conductor fractures and to decrease the incidence and number of inappropriate shocks.

OBJECTIVE

The purpose of this prospective single-center study was to assess the effectiveness of the LIA algorithm for warning patients of an impending Sprint Fidelis P/S conductor fracture and for decreasing the incidence and number of inappropriate shocks.

METHODS

The study population included all patients who had Sprint Fidelis leads and Medtronic ICD pulse generators that were implanted and followed at the Minneapolis Heart Institute. Patients were evaluated in the clinic every 3 to 4 months or by remote monitoring using the Medtronic CareLink system. When the LIA algorithm was released in August 2008, the RAMware was downloaded to the pulse generator of all patients with the Sprint Fidelis lead. Patients and family members received educational materials and were given a demonstration of the audible alerts.

RESULTS

Between October 2004 and January 2010, 52 (11.3%) of 461 Sprint Fidelis leads failed in the study population. Inappropriate shocks were the first sign of lead failure in 18 (69%) of the 26 patients who did not have the LIA compared to 4 (17%) of 23 patients who had the LIA (P = .0004). Patients who experienced inappropriate shocks without the LIA received an average of 13.2 +/- 13.6 inappropriate shocks (range 2-54) versus 3.0 +/- 2.0 inappropriate shocks (range 2-6) in patients who had the LIA (P = .017). The audible alert was effective in 70% (16/23) and 35% (6/17) of patients with and without the LIA, respectively, whose alerts were programmed ON (P = .053). Overall, 8 (32%) of 25 patients whose audible alerts were triggered did not immediately hear or recognize the tone.

CONCLUSION

The LIA appears to be an effective method for detecting most Sprint Fidelis lead fractures and for decreasing the incidence and number of inappropriate shocks. However, a better method for alerting patients and caregivers is needed.

摘要

背景

美敦力 Sprint Fidelis 高电压植入式心律转复除颤器 (ICD) 导联易发生断裂。大多数断裂涉及起搏感知 (P/S) 导线,可能导致多次不必要的电击。美敦力导联完整性警报 (LIA) 算法旨在提高对瞬态 P/S 导线断裂的早期检测,并降低不必要电击的发生率和数量。

目的

本前瞻性单中心研究的目的是评估 LIA 算法对警告 Sprint Fidelis P/S 导线即将发生断裂的患者的有效性,并降低不必要电击的发生率和数量。

方法

研究人群包括在明尼苏达心脏研究所植入和随访的所有具有 Sprint Fidelis 导联和美敦力 ICD 脉冲发生器的患者。患者每 3 至 4 个月在诊所接受评估,或使用美敦力 CareLink 系统进行远程监测。2008 年 8 月发布 LIA 算法后,将 RAMware 下载到所有具有 Sprint Fidelis 导联的脉冲发生器中。患者及其家属接受了教育材料,并演示了可听警报。

结果

在 2004 年 10 月至 2010 年 1 月期间,研究人群中 52 个 (11.3%) 的 461 个 Sprint Fidelis 导联发生故障。在没有 LIA 的 26 名患者中,电击是导线故障的第一个信号,占 18 名患者 (69%),而在有 LIA 的 23 名患者中,占 4 名患者 (17%) (P =.0004)。没有 LIA 的经历电击的患者平均经历了 13.2 +/- 13.6 次不必要电击 (范围 2-54),而有 LIA 的患者经历了 3.0 +/- 2.0 次不必要电击 (范围 2-6) (P =.017)。在警报设置为 ON 的患者中,有 LIA 的患者中 70% (16/23) 和没有 LIA 的患者中 35% (6/17) 的可听警报有效 (P =.053)。总的来说,在其可听警报触发的 25 名患者中,有 8 名 (32%) 患者并未立即听到或识别出警报声。

结论

LIA 似乎是一种检测大多数 Sprint Fidelis 导联断裂并降低不必要电击发生率和数量的有效方法。但是,需要一种更好的方法来提醒患者和护理人员。